Infusion pumps recalled over hard limit messages blocking piggyback flush programming
ICU Medical is recalling Plum Duo Precision IV Pumps because clinicians may receive an Upper or Lower Hard Limit violation message that prevents a piggyback flush from being programmed.
The agency’s classification
FDA (Food): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: Scored 2 because the record's stated consequence is that a flush cannot be programmed, with no patient harm or reported injury.
Plain-English summary
The ICU Medical Plum Duo Precision IV Pump, list numbers 400020403 (US) and 400021003 (Philippines), is being recalled by ICU Medical, Inc.
The record states that Plum Solo and Duo infusion pumps include a feature that, when selected, automatically flushes the downstream line after a piggyback therapy, with the programmed flush volume delivered from the primary line container at the piggyback therapy rate once the piggyback therapy is complete. The record states that in certain cases the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) violation message when programming a piggyback flush, which prevents the flush from being programmed.
The record does not state a quantity. It covers list number 400020403 (US) on software version 1.2.2 and list number 400021003 (Philippines) on software version 1.2.0. Distribution was worldwide, covering the United States nationwide and the Philippines. The FDA has classified the action as Class II.
The stated consequence in the enforcement record is that the flush cannot be programmed, and no injuries are reported. Facilities should follow ICU Medical's instructions.
The recalled product
- Product
- Mini Gaceniga Authentic Cuban Loaf Cake Raisins (Net Wt. 12 oz (340g)) packaged in wax paper, placed in clear bag and packed in cardboard carton. 12/ 12 oz per master case.
- Manufacturer
- CATAO MARKET LLC
- Category
- Medical Device — Infusion pump
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Product identified by Best By dates (12/18/2025 through 03/18/2026)
Distribution
Distributed in 1 state:
Part of a larger recall action
This product is one of 4 recalled by CATAO MARKET LLC under a single FDA (Food) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · CATAO MARKET LLC
See all →- ModerateClassic Cuban loaf cake recalled over undeclared FD&C Yellow 5
FDA (Food) · 2026-01-28
- HighRayStation 10B treatment planning recalled over dose that fails to invalidate
FDA (Food) · 2026-01-28
- HighCustom manifold kits recalled over inflation handles that may detach
FDA (Food) · 2026-01-28
Same hazard · software fault
See all →- SevereWinnebago Ekko Motorhomes Recalled Over Generator Software Fault
NHTSA · 2026-07-30
- HighVolkswagen Jetta and Taos instrument panels may not display warnings
NHTSA · 2026-03-25
- HighWelch Allyn vision screener display may flicker inducing photosensitive seizures
FDA (Devices) · 2026-03-25
- HighFord and Lincoln image module resets can disable rearview camera
NHTSA · 2026-03-18
- HighHonda Prologue and Acura ZDX instrument and camera displays may fail
NHTSA · 2026-02-26
Similar recalls
- ModeratePlum Solo pumps recalled over hard limit messages blocking flush programming
FDA (Devices) · 2026-01-28
- ModeratePlum Duo infusion pumps recalled over blocked piggyback flush programming
FDA (Devices) · 2026-01-28
- HighICU Medical Plum Duo Infusion Pump Software Dose Limit Issue
FDA (Devices) · 2025-06-25
- HighPlum Duo Precision IV pumps may not display soft limit alerts
FDA (Devices) · 2026-06-17
- HighICU Medical Plum Solos Infusion Pump Software Safety Defect Recall
FDA (Devices) · 2025-06-25
- HighPlum Duo infusion pumps may not display soft limit alerts
FDA (Devices) · 2026-06-17