238 recalls in our database name Stryker Endoscopy as the manufacturing or recalling firm.

What the numbers show

Total
238
Critical
0
Severe
2
Most recent
2016-05-04

Of the 238 recalls from Stryker Endoscopy we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 2 (1%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

151–175 of 238

  • ModerateFDA (Devices)·Z-0628-2015·2014-12-24

    Stryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Allis Forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ALLIS FORCEPS, P/N 0250080240. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0673-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 units of atraumatic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ATRAUMATIC FORCEPS, DOUBLE ACTION, P/N 0250080311. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0771-2015·2014-12-24

    Stryker Endoscopy Recalls 3mm Biopsy Punch Instruments Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of 3mm Biopsy Punch instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, BIOPSY PUNCH, 20CM, P/N 0250282013 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0618-2015·2014-12-24

    Stryker Laparoscopic Manual Instruments Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, LUERLOCK, MALE, 6MM TUBING, P/N 0250080207. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0748-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0708-2015·2014-12-24

    Stryker Endoscopy Recalls 5mm Insert Biopsy Forceps Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of 5mm Insert Biopsy Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT BIOPSY FORCEPS, SPOON W/PIN, 45CM, P/N 0250080581 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0664-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Obturator Packages Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of 11mm Obturator packages because the instructions for use do not support specific sterilization methods.

    Product
    PKG, OBTURATOR, 11MM, FOR HASSON CANNULA, P/N 0250080158. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0663-2015·2014-12-24

    Stryker Endoscopy Recalls 11mm Trocars Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of 11mm trocars because the instructions for use do not support specific sterilization methods.

    Product
    PKG, TROCAR, 11MM, PYRAMIDAL TIP, P/N 0250080156. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0703-2015·2014-12-24

    Stryker Endoscopy Recalls 5mm Metzenbaum Scissors Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of 5mm Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, ENDO METZENBAUM SCISSORS, CURVED 45CM, P/N 0250080575 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0637-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 Mixter Forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MIXTER FORCEPS, P/N 0250080249. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0693-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled Due to Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 units of bipolar forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BIPOLAR FORCEPS, PADDLE, 45CM, P/N 0250080351. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0766-2015·2014-12-24

    Stryker Laparoscopic Grasper Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic graspers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, BLUNT WITH SPOON, GRASPER, 20CM, P/N 0250282008 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0621-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080230. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0790-2015·2014-12-24

    Stryker Dorsey Grasper Recall Due to Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 Dorsey Grasper instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, DORSEY GRASPER, 29CM, P/N 0250282097 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0757-2015·2014-12-24

    Stryker Laparoscopic Manual Instruments Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM PEEK HANDLE, 45CM, P/N 0250181164 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0720-2015·2014-12-24

    Stryker Endoscopy Recalls 5mm PEEK Monopolar Handles Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of 5mm PEEK Monopolar Handles because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM PEEK MONOPOLAR HANDLE, P/N 0250080617 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0599-2015·2014-12-24

    Stryker Laparoscopic Knot Pushers Recalled for Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic knot pushers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, KNOT PUSHER 5MM OPEN, P/N 0250080132 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0735-2015·2014-12-24

    Stryker Endoscopy Recalls 5mm Right Angle Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of 5mm Right Angle Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, RIGHT ANGLE FORCEP, 45CM P/N 0250080729 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0675-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BLUNT FORCEPS, BLUNT, SERRATED, SPOON, P/N 0250080313. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0652-2015·2014-12-24

    Stryker Maryland Forceps Recalled Due to Sterilization Parameter Errors

    Stryker Endoscopy is recalling Maryland Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MARYLAND FORCEPS, CURVED LEFT, P/N 0250080282. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0730-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ASPIRATION NEEDLE, 33CM, P/N 0250080702 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0737-2015·2014-12-24

    Stryker Endoscopy Recalls Fenestrated Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 fenestrated forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM, 45CM, INSERT, FENESTRATED FORCEPS, P/N 0250080732 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0792-2015·2014-12-24

    Stryker Endoscopy Recalls 180,573 Laparoscopic Graspers Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 laparoscopic atraumatic graspers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, ATRAUMATIC GRASPER, DOUBLE ACTION, 20CM, P/N 0250282100 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0776-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Scissors Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, PERITONEAL SCISSORS, 29CM, P/N 0250282048 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0611-2015·2014-12-24

    Stryker Endoscopy Recalls Fixable Cone for 11mm Hasson Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 units of the Fixable Cone for 11mm Hasson because the instructions for use do not support specific sterilization methods.

    Product
    PKG, FIXABLE CONE FOR 11MM HASSON, P/N 0250080180. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide