Stryker Endoscopy Recalls Laparoscopic Obturator Packages Due to Sterilization Labeling Error
Stryker Endoscopy is recalling 180,573 units of 11mm Obturator packages because the instructions for use do not support specific sterilization methods.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II and involves a labeling error regarding sterilization methods without reported injuries or illnesses, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Stryker Endoscopy is recalling 180,573 units of PKG, OBTURATOR, 11MM, FOR HASSON CANNULA, P/N 0250080158. These laparoscopic manual instruments are intended for cutting, holding, grasping, and manipulating tissue and surgical auxiliaries. The recall covers units distributed nationwide in the United States and internationally to numerous countries including Canada, Mexico, Germany, Japan, and the United Kingdom.
The recall was initiated because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods. This is a labeling and documentation issue regarding the approved sterilization protocols for the device.
Healthcare facilities and distributors should stop using the affected products and contact Stryker Endoscopy for further instructions. The source text does not report any illnesses, injuries, or deaths associated with this recall.
The recalled product
- Product
- PKG, OBTURATOR, 11MM, FOR HASSON CANNULA, P/N 0250080158. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
- Manufacturer
- Stryker Endoscopy
- Affected units
- 180,573
Distribution
Part of a larger recall action
This product is one of 207 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 207 products in this recall →Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · mis labeling
See all →- SevereImperio Distribution Recalls Beef Jerky Produced Without Federal Inspection
USDA FSIS · 2026-10-24
- SevereInstrumentation Laboratory Recalls HemoslL AcuStar HIT-lgG In Vitro Diagnostic Reagents
FDA (Devices) · 2026-10-21
- LowMagellan LeadCare II Blood Lead Test Kit Recall for Labeling Error
FDA (Devices) · 2026-10-21
- ModerateIn2bones Recalls Partially Threaded Bone Screws Due to Labeling Error
FDA (Devices) · 2026-10-21
- ModerateGore Viabil Short Wire Biliary Endoprosthesis Recalled for Incorrect Length Labeling
FDA (Devices) · 2026-10-21
Similar recalls
- HighAesculap Hasson trocar recalled due to sterility packaging concerns
FDA (Devices) · 2024-03-13
- ModerateStryker Arthroscope and Hardware Set Tray Recall for Sterilization Labeling
FDA (Devices) · 2012-07-18
- ModerateStryker Recalls Mispackaged Orthopedic Shoulder Trial Instruments
FDA (Devices) · 2015-08-19
- ModerateStryker Irrigator Reusable Tips Recalled for Incorrect Sterilization Instructions
FDA (Devices) · 2013-05-15
- ModerateStryker Irrigator Tips Recalled for Incorrect Sterilization Temperature Instructions
FDA (Devices) · 2013-05-15
- HighStryker Recalls RF Cannulas Due to Packaging Error
FDA (Devices) · 2012-08-08