Stryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
Stryker Endoscopy is recalling 180,573 Allis Forceps units because the instructions for use do not support specific sterilization methods.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the source text does not report any illnesses or injuries, fitting the criteria for a moderate severity rating involving procedural or labeling issues without immediate harm.
Plain-English summary
Stryker Endoscopy is recalling 180,573 units of PKG, ALLIS FORCEPS (P/N 0250080240). These laparoscopic manual instruments are intended for cutting, holding, grasping, and manipulating tissue, organs, and surgical auxiliaries. The devices are distributed worldwide, including nationwide in the United States and in numerous international locations.
The recall is issued because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods. The source text does not report any specific illnesses, injuries, or deaths associated with this recall.
Healthcare facilities and distributors should stop using the affected devices and contact Stryker Endoscopy for further instructions regarding the return or replacement of the units.
The recalled product
- Product
- PKG, ALLIS FORCEPS, P/N 0250080240. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them…
- Manufacturer
- Stryker Endoscopy
- Affected units
- 180,573
Distribution
Part of a larger recall action
This product is one of 207 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 207 products in this recall →Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Alligator Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · sterilization instruction error
See all →- ModerateOmniTract Surgical Retractor System sterilization instructions require extended dry time
FDA (Devices) · 2026-07-15
- HighStryker Johann Forceps Recalled Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Similar recalls
- ModerateStryker Arthroscope and Hardware Set Tray Recall for Sterilization Labeling
FDA (Devices) · 2012-07-18
- ModerateStryker 12-Instrument Sterilization Tray Recalled for Validation Testing Gaps
FDA (Devices) · 2014-05-14
- HighSignature Laparoscopic Allis Clamp Grasper with Rotating Shaft Recalled
FDA (Devices) · 2025-06-11
- HighAlcon Single-Use Iris Bipolar Forceps Recalled for Lack of Sterilization
FDA (Devices) · 2014-12-17
- ModerateStingray Surgical Recalls Bipolar Forceps Due to Uncleared Instructions
FDA (Devices) · 2012-07-25
- ModerateInstrumed Reusable Laparoscopic Instruments Recalled for Updated Safety Instructions
FDA (Devices) · 2014-10-08