Stryker 12-Instrument Sterilization Tray Recalled for Validation Testing Gaps
Stryker Endoscopy is recalling 2,591 12-Instrument Sterilization Trays because gravity sterilization parameters in the reprocessing guide did not fully consider all worst-case scenarios during validation testing.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a gap in validation testing for sterilization parameters, which is a theoretical risk of inadequate sterilization rather than a reported injury or confirmed contamination event, fitting the criteria for a moderate severity precautionary recall.
Plain-English summary
Stryker Endoscopy is recalling 2,591 units of the 12-Instrument Sterilization Tray (Model 0242000012). The product is used to hold and protect surgical instruments during the sterilization process, specifically for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid with perforations to allow sterilizing agents to reach the devices inside, featuring silicon device holders to secure them.
The recall was initiated because the gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing. This issue affects the specific lot and serial numbers listed in the agency's recall record. The product was distributed worldwide.
Healthcare facilities and users should stop using the affected trays and contact Stryker Endoscopy for instructions on return or replacement. The recall is classified as Class II by the FDA, indicating a reasonable probability of temporary or medically reversible health consequences.
The recalled product
- Product
- 12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and…
- Manufacturer
- Stryker Endoscopy
- Affected units
- 2,591
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model number: 0242000012
- Lot/serial numbers: 021464-2
- 39824-1-1
- 4014
- 643893-1-4
- 27419
- 42901-1-1
- 35728-2-1
- 48639-1-8
- 021464-3
- 40288-1-1
- 43020
- 43893-1-6
- 27701
- 43104-1-1
- 36175-1-1
- 48956-1-1
- 23027
- 40629-3-
- 1
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · inadequate sterilization
See all →- HighBD Medical Extension Sets Recalled for Potential Sterilization Defects
FDA (Devices) · 2026-08-19
- HighOlympus Toomey Evacuator 30cc medical device sterilization parameters recalled
FDA (Devices) · 2026-07-22
- HighKarl Storz Ureteroscope Recall Due to Unapproved Reprocessing Instructions
FDA (Devices) · 2025-02-05
- HighMedline procedure kits recalled for surgical masks failing sterilization tests
FDA (Devices) · 2025-02-05
- ModerateKarl Storz Hopkins Telescope 6 endoscope recalled for unapproved reprocessing instructions
FDA (Devices) · 2024-10-02
Similar recalls
- ModerateStryker Arthroscope and Hardware Set Tray Recall for Sterilization Labeling
FDA (Devices) · 2012-07-18
- HighStryker Shaver Handpiece Sterilization Tray Recalled for Validation Deficiencies
FDA (Devices) · 2014-07-02
- ModerateStryker Laparoscopic Trocar Recall Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls 11mm Trocars Due to Sterilization Documentation Issues
FDA (Devices) · 2014-12-24
- ModerateStryker Laparoscopic Manual Instruments Recalled for Sterilization Parameter Errors
FDA (Devices) · 2014-12-24