Stryker Endoscopy Recalls Fixable Cone for 11mm Hasson Due to Sterilization Labeling
Stryker Endoscopy is recalling 180,573 units of the Fixable Cone for 11mm Hasson because the instructions for use do not support specific sterilization methods.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the source text explicitly states no illnesses or injuries are reported, fitting the criteria for a moderate severity score involving labeling or documentation issues.
Plain-English summary
Stryker Endoscopy is recalling 180,573 units of the PKG, FIXABLE CONE FOR 11MM HASSON (P/N 0250080180). These laparoscopic manual instruments are intended for cutting, holding, grasping, and manipulating tissue and organs during surgery. The recall covers units distributed worldwide, including the United States and numerous international locations.
The recall was initiated because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods. This is a Class II recall, indicating a situation that could cause temporary or medically reversible health consequences.
Healthcare facilities and distributors should stop using the affected devices and contact Stryker Endoscopy for further instructions. The source text does not report any illnesses or injuries associated with this recall.
The recalled product
- Product
- PKG, FIXABLE CONE FOR 11MM HASSON, P/N 0250080180. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
- Manufacturer
- Stryker Endoscopy
- Affected units
- 180,573
Distribution
Part of a larger recall action
This product is one of 207 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 207 products in this recall →Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · mis labeling
See all →- SevereWhite Vein Advanced Alkaloids Chewable Tablets Recall for Excess Active Ingredient
FDA (Food) · 2026-10-14
- HighJ K Patisserie Chocolate Chip Cookies Recalled for Undeclared Soy
FDA (Food) · 2026-10-14
- HighFreshPoint Recalls Cranberry Chicken Salad Wedges for Undeclared Egg Allergen
FDA (Food) · 2026-10-14
- HighJ K Patisserie & Bakery LLC Recalls Croissants for Undeclared Egg
FDA (Food) · 2026-10-14
- LowMindray ePM Series Patient Monitor Manuals Recalled for Alarm Clarification
FDA (Devices) · 2026-10-14
Similar recalls
- HighStryker 1.2mm Cross Cut Fissure Recalled for Endotoxin Levels
FDA (Devices) · 2024-10-23
- HighAesculap Hasson trocar recalled due to sterility packaging concerns
FDA (Devices) · 2024-03-13
- ModerateStryker 12-Instrument Sterilization Tray Recalled for Validation Testing Gaps
FDA (Devices) · 2014-05-14
- ModerateStryker Irrigator Reusable Tips Recalled for Incorrect Sterilization Instructions
FDA (Devices) · 2013-05-15
- HighStryker Hoffmann II Connecting Rod Micro Recalled Due to Potential Cracking
FDA (Devices) · 2026-08-05
- ModerateStryker Arthroscope and Hardware Set Tray Recall for Sterilization Labeling
FDA (Devices) · 2012-07-18