The Recall Desk

State

Utah product recalls

73,827 recalls have nationwide distribution and so reach Utah. 4,916 additional recalls listed Utah specifically in their distribution scope.

What the numbers show

Total
4,916
Critical
524
Severe
1,351
Most recent
2026-11-04

Of the 4,916 recalls naming Utah we have scored against our rubric, 524 (11%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,351 (27%) are Severe.

These figures cover only the 4,916 recalls whose distribution notice names Utah specifically. The 73,827 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

47876–47900 of 78743

  • HighFDA (Devices)·Z-1144-2017·2017-02-15

    Physio-Control Recalls LIFEPAK 15 Defibrillators with Erroneous CO2 Readings

    Physio-Control is recalling LIFEPAK 15 Monitor/Defibrillators with End-Tidal CO2 options because the CO2 reading may intermittently display an incorrect value of XXX.

    Product
    LIFEPAK 15 Monitor/Defibrillator with End-Tidal CO2 option. Intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1149-2017·2017-02-15

    Stryker Leibinger Recalls Surgical Drainage Accessories Due to Sterility Labeling Error

    Stryker Leibinger is recalling QUINTUBE Monitor Packs and SPG Bulb Drain Systems because non-sterile products were incorrectly labeled as sterile.

    Product
    QUINTUBE Monitor Pack (5 tubes/pk, 24 pk/case) (Non-Sterile) SPG Bulb Drain System, 4mm Flat w/Center Holes (Sterile) Accessories to TLS Surgical Drainage System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1142-2017·2017-02-15

    Merge Healthcare Recalls Eye Station Imaging Devices Due to Capture Failures

    Merge Healthcare is recalling 182 Eye Station capture stations because they may fail to capture eye images as intended, posing a risk to patient health.

    Product
    Merge Eye Station. f/k/a: DFC-1024 & DFC-512 Digital Imaging System, WinStation, and WinStation Retinal Imager distributed by Ophthalmic Imaging Systems (OIS). Model number 136 T1700 Capture Stations & 46 T5810 Capture Stations.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1148-2017·2017-02-15

    AtriCure Recalls AtriClip LAA Exclusion System Due to Jaw Breakage

    AtriCure is recalling 464 AtriClip LAA Exclusion System devices because a complaint was received regarding the PRO2 jaw breaking prior to surgery.

    Product
    AtriClip" LAA Exclusion System with preloaded Gillinov-Cosgrove" Clip - A single PRO2 device is packaged in a conventional thermoformed polyethylene terephthalate glycol (PETG) tray with a Tyvek¿ lid. A single-walled non-corrugated box completes the single unit package. Six (6)…
    Category
    Medical Device
    Distribution
    31 states
  • HighFDA (Devices)·Z-1176-2017·2017-02-15

    Merge PACS Software Recalled for Potential Incorrect Patient Image Display

    Merge Healthcare is recalling Merge PACS software version 7.0.1 because it may display incorrect patient images, potentially delaying diagnosis or treatment.

    Product
    Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-1177-2017·2017-02-15

    Teleflex Recalls Sheridan Cuffed Tracheal Tubes Due to Labeling Inconsistency

    Teleflex Medical is recalling Sheridan cuffed tracheal tubes because some units are labeled with incorrect sizes, potentially leading to improper airway management.

    Product
    Sheridan/CF, Cuffed Tracheal Tube, Catalog # 5-10114 and Sheridan/HVT Cuffed Tracheal Tube, Catalog # 5-10315 A tracheal tube is inserted into a patients mouth or nose for airway management
    Category
    Medical Device
    Distribution
    38 states
  • HighFDA (Food)·F-1440-2017·2017-02-15

    Dannon Dairy Drink Recalled for Incorrect Walnut Allergen Labeling

    Dannon is recalling specific bottles of Dairy Drink Nut and Cereal because the label incorrectly lists walnut as an ingredient and allergen, while the product actually contains pecan.

    Product
    DANONE¿ DAIRY DRINK NUT AND CEREAL 7.7 FL OZ (229mL) bottles, UPC 36632 01421
    Category
    Food
    Distribution
    18 states
  • ModerateFDA (Devices)·Z-1143-2017·2017-02-15

    Physio-Control Recalls LIFEPAK 12 Defibrillators with Erratic CO2 Readings

    Physio-Control is recalling LIFEPAK 12 Defibrillators because the End-Tidal CO2 reading may intermittently display an invalid value of XXX during operation.

    Product
    LIFEPAK 12 Defibrillator/Monitor with End-Tidal CO2 option. The device is used in various areas of the hospital such as critical areas (emergency departments, critical care, operating room, etc.) and general duty floors (e.g. medical/surgical). The device is also used…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1150-2017·2017-02-15

    Stryker Leibinger Recalls Non-Sterile Medpor Implant Sizer Sets Due to Labeling Error

    Stryker Leibinger is recalling specific non-sterile Medpor implant sizer sets because they were incorrectly labeled as sterile. The recall covers 619 units distributed in the US and several international countries.

    Product
    9805 Orbital Volume Sizer Set with Tray (Non-sterile) 9951 Design M Malar Sizer Set (Silicone, Non-Sterile) 9952 Extended Contoured, Malar Sizer Set (Silicone, Non-Sterile) 85000 Petite Nasal Dorsum Sizer Set (Silicone, Non-Sterile) Accessories to Medpor Implants
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Drugs)·D-0469-2017·2017-02-15

    VistaPharm Recalls Potassium Chloride Oral Solution Due to Cup Leakage

    VistaPharm is recalling 13,609 cases of Potassium Chloride Oral Solution because the unit dose cups may leak at the seal.

    Product
    Potassium Chloride Oral Solution, USP, 10%, 20 mEq per 15 mL, packaged in 15 mL unit dose cups (NDC 66689-047-01), 50 unit dose cups per case (NDC 66689-047-50), Rx only, Xact Dose, Manufactured by VistaPharm, Largo, FL 33771.
    Category
    Drug
    Distribution
    36 states
  • ModerateFDA (Devices)·Z-1140-2017·2017-02-15

    Siemens Dimension Ammonia Reagent Cartridges Recalled for Calibration Issues

    Siemens Healthcare Diagnostics is recalling specific lots of Dimension Ammonia Flex reagent cartridges because they may not meet the 60-day calibration interval claim, potentially leading to abnormal assay results.

    Product
    Dimension¿ Ammonia Flex¿ reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in human plasma on the Dimension clinical chemistry system. Ammonia measurements are used in the diagnosis and treatment of…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1141-2017·2017-02-15

    Siemens Dimension Vista Ammonia Flex Reagent Cartridge Recall

    Siemens is recalling specific lots of Dimension Vista Ammonia Flex reagent cartridges because they may not meet the 60-day calibration interval claim, potentially resulting in abnormal assay results.

    Product
    Dimension Vista¿ Ammonia Flex¿ reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in human plasma on the Dimension clinical chemistry system. Ammonia measurements are used in the diagnosis and treatment…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1181-2017·2017-02-15

    Ortho VITROS FS Calibrator 1 Recalled Due to Calibration Failure

    Ortho-Clinical Diagnostics is recalling VITROS FS Calibrator 1, Lot 91461, because it cannot calibrate dLDL reagents when used with specific Assay Data Disk versions.

    Product
    VITROS Chemistry Products FS Calibrator 1, REF 680 1873 12 x 3 mL, Rx ONLY, IVD; used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS/4600 Chemistry and integrated Systems. --- Ortho-Clinical Diagnostics, Inc. Rochester, NY 14626…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1139-2017·2017-02-15

    Encore Medical Recalls Empowr PS Knee Tibial Inserts Due to Missing Feature

    Encore Medical is recalling 24 Empowr PS Knee Tibial Inserts because they are missing the impaction slot feature.

    Product
    Empowr PS Knee Tibial Insert; Model 343-14-709 The EMPOWR PS tibial insert are made from Highly Cross-Linked Vitamin E (HXL VE) UHMWPE. The tibial insert implants are intended to mate with tibial base implants.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-1179-2017·2017-02-15

    Ortho-Clinical VITROS UPRO Slides Recalled for Variable Urine Protein Results

    Ortho-Clinical Diagnostics is recalling VITROS UPRO Slides because they may produce variable urine protein results.

    Product
    VITROS Chemistry Products UPRO Slides, UPN 10758750000593, Product Usage: Quantitative measurement of total protein concentration in urine.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Devices)·Z-1146-2017·2017-02-15

    Thayer MiniSpacer MDI Adapter Recalled for Incorrect Labeling

    Thayer Medical Corporation is recalling 2,550 MiniSpacer MDI adapters because the insert and case labels incorrectly state the connection size.

    Product
    Thayer MiniSpacer Dual Spray MDI Adapter Counter Incrementing Actuator with 15mm O.D./I.D. Connections, Reference no. 1543A. The 15 mm Dual Spray MiniSpacer with Counter Incrementing Actuator is a non-sterile, single-use item, intended to dispense aerosolized medication into a…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • LowFDA (Drugs)·D-0470-2017·2017-02-15

    Aurobindo Recalls Discolored Pantoprazole Sodium Injection Vials

    Aurobindo Pharma USA Inc is recalling 29,800 vials of Pantoprazole Sodium for Injection due to discoloration. The affected vials contain powder with a yellowish-brownish appearance.

    Product
    PANTOPRAZOLE SODIUM — PANTOPRAZOLE SODIUM (PANTOPRAZOLE SODIUM)
    Category
    Drug
    Distribution
    Distributed nationwide
  • LowFDA (Drugs)·D-0471-2017·2017-02-15

    Apotex Recalls Aripiprazole Tablets Due to Potential Potency Issues

    Apotex Inc. is recalling 67,946 bottles of 2 mg Aripiprazole tablets because the product may not meet specifications throughout its shelf life.

    Product
    Aripiprazole Tablets, 2 mg, 30-count bottle, Rx only, Manufactured by: Apotex Inc., Toronto, Ontario, Canada, M9L 1T9; Manufactured for: Apotex Corp., Weston, Florida 33326, NDC 60505-3075-3.
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereNHTSA·17V097000·2017-02-14

    Chrysler Recalls 2014-2017 AWD Vehicles for Driveshaft Bolt Issue

    Chrysler is recalling 2014-2017 Dodge Charger and Chrysler 300 AWD vehicles because front driveshaft bolts may loosen, potentially causing a loss of motive power and increasing crash risk.

    Product
    CHRYSLER — CHRYSLER, DODGE 300, CHARGER
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·17V096000·2017-02-14

    Thor Motor Coach Recalls 2014-2016 Motorhomes for Wiper Defect

    Thor Motor Coach is recalling 2,483 motorhomes because improperly installed windshield wipers may fail, reducing visibility and increasing crash risk.

    Product
    THOR MOTOR COACH — THOR MOTOR COACH, THOR TUSCANY XTE, TUSCANY, PALAZZO
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • SevereNHTSA·17V095000·2017-02-13

    Forest River Recalls 2017 Palomino Columbus RVs Due to Propane Hose Defect

    Forest River is recalling 2017 Palomino Columbus recreational vehicles equipped with an optional generator package because the liquid propane supply hose may be too small, posing a fire or explosion risk.

    Product
    PALOMINO — PALOMINO COLUMBUS
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighCPSC·17087·2017-02-13

    Target Recalls Threshold Patio Benches Due to Fall Hazard

    Target is recalling about 1,300 Threshold patio benches because they can collapse while in use, posing a fall hazard.

    Product
    Threshold patio benches
    Category
    Consumer Product
    Distribution
    Distributed nationwide