  • SevereFDA (Drugs)·D-0653-2026·2026-07-08

    IMULDOSA injection recalled due to lack of sterility assurance

    Accord BioPharma is recalling IMULDOSA (ustekinumab-srlf) Injection due to lack of assurance of sterility. The affected lot was distributed nationwide.

    Product
    IMULDOSA — IMULDOSA (USTEKINUMAB-SRLF)
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-2550-2026·2026-07-08

    Hillrom VOLARA System patient circuits recalled for nebulizer leakage risk

    Baxter Healthcare is recalling Hillrom VOLARA System patient circuits due to reports of air and medication leakage from the nebulizer cup, which can reduce oxygen levels and prevent proper medication delivery.

    Product
    Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Drugs)·D-0643-2026·2026-07-08

    ChloraPrep One-Step Applicators Recalled for Lack of Sterility Assurance

    CareFusion 213, LLC is recalling ChloraPrep One-Step applicators that were intended for sterilization but were instead distributed unsterilized to customers outside the intended distribution channel.

    Product
    CHLORAPREP ONE-STEP — CHLORAPREP ONE-STEP (CHLORHEXIDINE GLUCONATE AND ISOPROPYL ALCOHOL)
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereFDA (Drugs)·D-0676-2026·2026-07-08

    Cinacalcet Hydrochloride Tablets Recalled for Elevated Impurity Levels

    Cipla USA, Inc. is recalling Cinacalcet Hydrochloride 90 mg tablets (Lot #5PB0563, expiring 4/30/2027) distributed nationwide due to the presence of N-nitroso-cinacalcet above acceptable daily intake levels, a manufacturing deviation.

    Product
    CINACALCET — CINACALCET (CINACALCET)
    Category
    Drug
    Distribution
    Distributed nationwide
  • SevereNHTSA·26V438000·2026-07-08

    BMW and Toyota vehicles recalled due to engine starter fire risk

    BMW of North America, LLC is recalling various BMW and Toyota Supra models because the engine starter may overheat and catch fire.

    Product
    TOYOTA — TOYOTA, BMW SUPRA, Z4, X4, X3
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2637-2026·2026-07-08

    CRE Wireguided 18-20mm 240cm medical devices recalled for sterile breach

    Boston Scientific Corporation is recalling CRE Wireguided 18-20mm 240cm devices worldwide due to a potential sterile breach of the pouches in which the devices are packaged.

    Product
    CRE Wireguided 18-20mm 240cm
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2608-2026·2026-07-08

    Medline Convenience Kits with Recalled Swan-Ganz Catheters

    Medline Industries is recalling convenience kits containing Swan-Ganz catheters due to defects in the proximal injectate lumen hub that may leak or break, risking infection, medication loss, and blood loss.

    Product
    Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2609-2026·2026-07-08

    Medline Convenience Kits Containing Recalled Swan-Ganz Catheters

    Medline is recalling convenience kits containing Swan-Ganz catheters due to leaks and breakage of the proximal injectate lumen hub, which may cause infection, medication loss, or blood loss.

    Product
    Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2627-2026·2026-07-08

    CRE Pro Wireguided 8-10mm 240cm Medical Device Sterile Pouch Breach

    Boston Scientific is recalling CRE Pro Wireguided 8-10mm 240cm devices worldwide due to a potential sterile breach of the product packaging pouches. The breach could compromise the sterility of the devices.

    Product
    CRE Pro Wireguided 8-10mm 240cm
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2612-2026·2026-07-08

    Medline Convenience Kits with Recalled Swan-Ganz Catheters

    Medline Convenience kits containing recalled Swan-Ganz catheters may leak or break at the proximal injectate lumen hub, risking infection, medication loss, and blood loss.

    Product
    Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2625-2026·2026-07-08

    IntelliCuff Portable Automatic Cuff Pressure Controller Recall

    Hamilton Medical AG is recalling the IntelliCuff portable automatic cuff pressure controller due to a reported "Cuff System Leakage" error that may cause the motor to pump continuously and fail to maintain user-set cuff pressure.

    Product
    IntelliCuff, Portable automatic cuff pressure controller REF: 951001
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2638-2026·2026-07-08

    Baxter Head Positioning Adapter May Loosen During Patient Positioning

    Baxter Healthcare is recalling an Adapter for Head Positioning (Model 1739994) used on operating tables because it may become loose and unintentionally move during patient positioning.

    Product
    Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2618-2026·2026-07-08

    Medacta Short Monobloc Stem Broaches Surgical Instruments Shoulder Implant

    Medacta USA is recalling Short Monobloc Stem Broaches used in shoulder replacement surgery due to positioning misalignment that may increase risk of fracture in the upper arm bone.

    Product
    Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2647-2026·2026-07-08

    RayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss

    RayCare 2024A SP4 oncology information system may lose data or fail to notify users when beam set delivery notes are modified simultaneously or updated during treatment sessions.

    Product
    RayCare 2024A SP4. Oncology Information Systems.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2611-2026·2026-07-08

    Medline Convenience Kits with Recalled Swan-Ganz Catheters Nationwide

    Medline is recalling convenience kits containing Swan-Ganz catheters with a defective proximal injectate lumen hub that may leak, break, or cause infection and blood loss.

    Product
    Medline Convenience kits containing recalled Swan-Ganz Catheters OPEN HEART DYNJ909090B OPEN HEART CDS QVH CDS982523M
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0655-2026·2026-07-08

    Prednisolone Acetate Ophthalmic Suspension Recalled for Foreign Substance

    Lupin Pharmaceuticals is recalling Prednisolone Acetate Ophthalmic Suspension, USP, 1% due to the presence of a foreign substance. Affected lot numbers were distributed nationwide.

    Product
    PREDNISOLONE ACETATE — PREDNISOLONE ACETATE (PREDNISOLONE ACETATE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighNHTSA·26V440000·2026-07-08

    Shyft Recalls Utilimaster Velocity Vehicles Due to Strobe Light Fire Risk

    Shyft Group is recalling 2016, 2019, and 2021 Utilimaster Velocity vehicles because water entering an improperly sealed strobe light may cause a short circuit and increase the risk of fire.

    Product
    UTILIMASTER — UTILIMASTER VELOCITY
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2626-2026·2026-07-08

    Boston Scientific CRE Pro Wireguided catheters recalled for packaging

    Boston Scientific is recalling CRE Pro Wireguided 6-8mm 240cm catheters worldwide due to a potential sterile breach in the packaging pouches, which could compromise device sterility.

    Product
    CRE Pro Wireguided 6-8mm 240cm
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2623-2026·2026-07-08

    GC Agar Base Culture Media Reduced Recovery of Neisseria gonorrhoeae

    Remel, Inc is recalling GC Agar Base culture media because some strains of Neisseria gonorrhoeae may not be recovered or may show reduced recovery on the product.

    Product
    GC Agar Base (150mm) 10/PK R04030; GC Agar Base (100mm) 10/PK R01460
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0661-2026·2026-07-08

    Cinacalcet Tablets Recalled for N-Nitroso Impurity Exceeding Limits

    Cipla USA, Inc. is recalling Cinacalcet Hydrochloride 30 mg tablets due to the presence of N-nitroso-cinacalcet above the acceptable daily intake. The affected product was distributed nationwide.

    Product
    CINACALCET — CINACALCET (CINACALCET)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighNHTSA·26V443000·2026-07-08

    Husqvarna and GasGas Motorcycles Recalled for Brake Caliper Cracks

    KTM North America is recalling certain 2022-2024 Husqvarna FE 501S, FE 350S, and 2024 GasGas ES 350 and ES 500 motorcycles due to a potential rear brake caliper crack that could reduce braking ability and increase crash risk.

    Product
    HUSQVARNA — HUSQVARNA, GASGAS FE 501S, ES 500, ES 350, FE 350S
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2613-2026·2026-07-08

    Medline Convenience Kits with Recalled Swan-Ganz Catheters

    Medline Convenience kits containing Swan-Ganz catheters recalled by Becton Dickinson are being recalled due to leaks and breakage of the proximal injectate lumen hub, which may result in infection, medication loss, or blood loss.

    Product
    Medline Convenience kits containing recalled Swan-Ganz Catheters CABG A CDS CDS982269K CDS982269L CDS982269M CDS982269N CARDIAC MAJOR DYNJ901328D DYNJ901328F HEART A DYNJ903465F DYNJ903465G HEART TRAY-VASCULAR DYNJ907857 DYNJ907857A DYNJ90
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2643-2026·2026-07-08

    ARTIS icono floor Medical Imaging System Detector Movement Loss Recall

    Siemens Medical Solutions is recalling 30 units of the ARTIS icono floor (Model 11327700) due to potential loss of detector movement capability when adjusted to maximum or minimum height limits.

    Product
    ARTIS icono floor. Model Number: 11327700.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2639-2026·2026-07-08

    Pilling Wecksorb Neurosponges Surgical Sponges Recalled for Endotoxin Contamination

    Teleflex LLC is recalling Pilling Wecksorb Neurosponges and Neurosponge Strips due to out-of-specification or potentially out-of-specification endotoxin levels. The cotton surgical sponges are used in brain and spine surgery.

    Product
    Brand Name: Pilling Wecksorb Product Name: Pilling Wecksorb Neurosponges (Cotton) Pilling¿ Wecksorb Neurosponge Strips (Cotton) Model/Catalog Number: 200100 200101 200102 200103 200104 200105 200106 200108 200110 200112 200120 200121 200123 200131 Software V
    Category
    Medical Device
    Distribution
    Distributed nationwide