  • SevereFDA (Devices)·Z-1546-2026·2026-03-18

    TRUE METRIX PRO manuals recalled as Class I over E-5 guidance

    Trividia Health is recalling owner's booklets and instructions for TRUE METRIX PRO blood glucose meters because the labeling did not give lay users adequate directions for the E-5 error code.

    Product
    Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1547-2026·2026-03-18

    TRUE METRIX AIR manuals recalled as Class I over E-5 guidance

    Trividia Health is recalling owner's booklets and instructions for TRUE METRIX AIR blood glucose meters because the labeling did not give lay users adequate directions for the E-5 error code.

    Product
    Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0548-2026·2026-03-18

    Vegetarian Chicken Flavor Ramen Express contains undeclared Yellow #5

    Palmetto Gourmet Foods Inc. is recalling Vegetarian Chicken Flavor Ramen Express because the product contains undeclared Yellow #5 food coloring, a major allergen for some consumers.

    Product
    VEGETARIAN CHICKEN FLAVOR RAMEN Express Net wt 3oz (85g) Ingredients: Enriched unbleached flour (wheat flour***) Palm oil***" Contains: Wheat, Soy. UPC 8 10037 81004 2 Produced by PALMETTO GOURMET FOODS, INC SALUDA, SC 29138
    Category
    Food
    Distribution
    32 states
  • HighFDA (Devices)·Z-1523-2026·2026-03-18

    Link Endo-Model plateau 15-0027/16 recalled over screw bushing fracture

    Waldemar Link is recalling Endo-Model Replacement Plateau item 15-0027/16 because the bushing on a plateau screw could detach from the screw shaft due to a longitudinal fracture.

    Product
    Endo-Model Replacement Plateau; Item Number: 15-0027/16;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1528-2026·2026-03-18

    GEM Premier 5000 PAK part 00055407504 may eject during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 PAK part 00055407504 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000 PAK; Part No. 00055407504.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1520-2026·2026-03-18

    Link Endo-Model plateau 15-8521/15 recalled over screw bushing fracture

    Waldemar Link is recalling Endo-Model Replacement Plateau item 15-8521/15 because the bushing on a plateau screw could detach from the screw shaft due to a longitudinal fracture.

    Product
    Endo-Model Replacement Plateau; Item Number: 15-8521/15;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1532-2026·2026-03-18

    GEM Premier 5000 PAK part 00055415004 may eject during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 PAK part 00055415004 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000 PAK; Part No. 00055415004.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1509-2026·2026-03-18

    Exactech Ergo modular impactor handle may be missing its cross-pin

    Exactech is recalling the Equinoxe Ergo Modular Impactor Handle, model 321-09-05, because the impactor handle may be missing a cross-pin.

    Product
    Brand Name: Equinoxe Ergo Product Name: Ergo Modular Impactor Handle Model/Catalog Number: 321-09-05
    Category
    Medical Device
    Distribution
    25 states
  • HighFDA (Devices)·Z-1512-2026·2026-03-18

    Link Endo-Model plateau 15-0027/15 recalled over screw bushing fracture

    Waldemar Link is recalling Endo-Model Replacement Plateau item 15-0027/15 because the bushing on a plateau screw could detach from the screw shaft due to a longitudinal fracture.

    Product
    Endo-Model Replacement Plateau; Item Number: 15-0027/15;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·26V161000·2026-03-18

    2026 Keystone Montana Fifth Wheel Recreational Trailers Electrical Wiring Recall

    Keystone is recalling certain 2026 Montana fifth wheel trailers because improperly secured 12-volt and 120-volt wiring may be pinched and damaged when the bedroom slide-out moves, increasing fire risk.

    Product
    KEYSTONE — KEYSTONE MONTANA
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1497-2026·2026-03-18

    IBA proton therapy system may not interrupt beam when gating fails

    Ion Beam Applications is recalling PROTEUS 235 proton therapy systems because a disabled interface means the safety system may not interrupt beam delivery if the external gating signal is lost.

    Product
    IBA Proton Therapy System - PROTEUS 235
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1522-2026·2026-03-18

    Link Endo-Model plateau 15-0027/12 recalled over screw bushing fracture

    Waldemar Link is recalling Endo-Model Replacement Plateau item 15-0027/12 because the bushing on a plateau screw could detach from the screw shaft due to a longitudinal fracture.

    Product
    Endo-Model Replacement Plateau; Item Number: 15-0027/12;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1554-2026·2026-03-18

    Artelon FlexBand Dynamic Matrix failed bacterial endotoxin testing

    International Life Sciences is recalling the Artelon FlexBand Dynamic Matrix, ref 31057, because the augmentation devices failed bacterial endotoxin testing.

    Product
    Artelon FlexBand Dynamic Matrix Ref: 31057
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1516-2026·2026-03-18

    Link Endo-Model plateau 15-2836/11 recalled over screw bushing fracture

    Waldemar Link is recalling Endo-Model Replacement Plateau item 15-2836/11 because the bushing on a plateau screw could detach from the screw shaft due to a longitudinal fracture.

    Product
    Endo-Model Replacement Plateau; Item Number: 15-2836/11;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1552-2026·2026-03-18

    Beckman Coulter DxC 700 AU analyzer may stop analyzing after errors

    Beckman Coulter is recalling DxC 700 AU analyzers because leaving the calibration monitor open beyond 15 minutes can trigger database errors that stop the system analyzing.

    Product
    DxC 700 AU, REF: B86444, B86446
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1511-2026·2026-03-18

    Link Endo-Model SL component 16-2840/05 recalled over bushing fracture

    Waldemar Link is recalling Endo-Model SL Connection Component item 16-2840/05 because the bushing on a plateau screw could detach from the screw shaft due to a longitudinal fracture.

    Product
    Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/05;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1525-2026·2026-03-18

    GEM Premier 5000 PAK part 00055360008 may eject during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 PAK part 00055360008 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000 PAK; Part No. 00055360008.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1534-2026·2026-03-18

    GEM Premier 5000 part 00055415008 may eject cartridges during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 part 00055415008 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000; Part No. 00055415008.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1535-2026·2026-03-18

    GEM Premier 5000 part 00055415010 may eject cartridges during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 part 00055415010 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000; Part No. 00055415010.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1555-2026·2026-03-18

    Artelon FlexBand Plus implants failed bacterial endotoxin testing

    International Life Sciences is recalling Artelon FlexBand Plus, refs 41054 and 41057, because the augmentation devices failed bacterial endotoxin testing.

    Product
    Artelon FlexBand Plus Ref: 41054 & 41057
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1513-2026·2026-03-18

    Link Endo-Model SL component 16-2840/02 recalled over bushing fracture

    Waldemar Link is recalling Endo-Model SL Connection Component item 16-2840/02 because the bushing on a plateau screw could detach from the screw shaft due to a longitudinal fracture.

    Product
    Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1531-2026·2026-03-18

    GEM Premier 5000 PAK part 00055407511 may eject during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 PAK part 00055407511 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000 PAK; Part No. 00055407511.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1540-2026·2026-03-18

    GEM Premier 5000 part 00055430011 may eject cartridges during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 part 00055430011 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000; Part No. 00055430011.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·26V165000·2026-03-18

    Ford and Lincoln image module resets can disable rearview camera

    Ford is recalling certain 2022-2025 Lincoln Navigator, Nautilus, Aviator and Explorer vehicles because the Image Processing Module A may reset unexpectedly, losing the rearview camera image and driver assistance features.

    Product
    LINCOLN — LINCOLN, FORD AVIATOR, EXPLORER, NAUTILUS, NAVIGATOR
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1529-2026·2026-03-18

    GEM Premier 5000 PAK part 00055407508 may eject during warm-up

    Instrumentation Laboratory is recalling GEM Premier 5000 PAK part 00055407508 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.

    Product
    GEM Premier 5000 PAK; Part No. 00055407508.
    Category
    Medical Device
    Distribution
    Distributed nationwide