SevereFDA (Devices)·Z-1547-2026·Announced 2026-03-18

TRUE METRIX AIR manuals recalled as Class I over E-5 guidance

Trividia Health is recalling owner's booklets and instructions for TRUE METRIX AIR blood glucose meters because the labeling did not give lay users adequate directions for the E-5 error code.

What this means for you

Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The rubric states an FDA Class I recall is never scored below 4, and this record carries a Class I classification with no reported deaths in the source.

Plain-English summary

Trividia Health, Inc. is recalling owner's booklets and instructions for use supplied with TRUE METRIX AIR blood glucose measuring devices, covering retailer-branded kits including Walgreens, Leader, CVS, GNP, Meijer, Publix, Discount Drug Mart, HEB, Sunmark, Rite Aid, Centerwell, Humana, Healthmart, Relion, Kroger and Foster and Thrive versions. About 3,678,026 units were distributed worldwide, covering the United States nationwide including Puerto Rico, and Mexico, the United Kingdom, Jamaica, Trinidad & Tobago and Australia.

The system labeling, in both the user manual and online, did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions do not clearly emphasise that an E-5 error code could represent a very high blood glucose level, and do not prominently convey the appropriate immediate actions such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose. The FDA has classified this recall as Class I, its most serious category.

Affected booklets include REA4TVH03 Rev 57 and prior, and REA4TVH35 Rev 51 and prior in English and Spanish.

Users can identify affected documentation by the part number and revision printed on it, and may contact Trividia Health, Inc. for further information.

The recalled product

Product
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-
Manufacturer
Trividia Health, Inc.
Hazard
  • inadequate-instructions
  • treatment-delay
  • labelling-error
  • class-i-recall

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Device Item/UDI-DI: REA4001-01/311917173894
  • REA4002-01/96295129267
  • REA4007-01/50428354766
  • REA4011-01/87701427398
  • REA4019-01/708820685659
  • REA4020-01/41415072787
  • REA4025-01/93351022784
  • REA4031-01/41220009169
  • REA4054-01/10939742445
  • REA4066-01/11822590051
  • REA4081-01/21292015015
  • REA4081-40/21292007164
  • REA4081-43/21292007584
  • REA4089-01/52569139042
  • REA4094-01/681131240154
  • REA4094-01RX/681131403214
  • REA4203-01/41260007354
  • REA4211-01/10939958617
  • REA4H01-01/21292007010
  • REA4H01-40/21292007447

Distribution

Distributed nationwide across the United States.