Severity 3 (High) recalls have a known hazard but no confirmed illness, injury, or property damage in the source text. The risk is theoretical-but-real: a pathogen pre-emptively pulled from shelves before anyone got sick, an allergen mislabeling on an uncommon allergen, a precautionary correction on a device that hasn't yet malfunctioned in the field. Most allergen recalls land here, as do most preventive food recalls based on routine sampling. You should still stop using the product, but you don't need to panic.

The full rubric is on the methodology page. Severity is computed deterministically from the agency’s own classification plus the source text — no editorial vibes.

17576–17600 of 50403

  • HighFDA (Devices)·Z-0857-2022·2022-04-13

    Signa 1.5T MRI System Recall: Potential Horizontal Image Reversal

    GE Healthcare is recalling 10 Signa 1.5T TwinSpeed MRI systems worldwide due to a potential defect where magnetic resonance images may be flipped left to right.

    Product
    Signa 1.5T TwinSpeed Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0871-2022·2022-04-13

    Draeger Babylog VN500 Ventilators Recalled for Cybersecurity Vulnerability

    Draeger Medical is recalling Babylog VN500 ventilators for neonatal and pediatric patients due to cybersecurity vulnerabilities. The devices are not equipped against potential cyber threats.

    Product
    Babylog VN500 Ventilator with Software 2.51.01 and Lower- intended for the ventilation neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight Catalog Number: 8417400
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·22E033000·2022-04-13

    Bendix Brake Light Switch Recall Due to Binding Connection

    Bendix is recalling brake light switches that may bind, causing lights to stay on or fail to illuminate, increasing crash risk.

    Product
    EXTERIOR LIGHTING:BRAKE LIGHTS:SWITCH recall
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0896-2022·2022-04-13

    C-Peptide ELISA Kit Recalled Due to Defective Standards

    DRG International recalls its C-Peptide ELISA kit (Catalog # EIA-1293) after the standards were found to have optical density below the required threshold. The defect may cause invalid test runs and delayed patient results.

    Product
    C-Peptide ELISA : IVD measurement of C-Peptide in serum, plasma (EDTA-, heparin- or citrate plasma) and urine. Catalog # EIA-1293
    Category
    Medical Device
    Distribution
    1 state
  • HighNHTSA·22V243000·2022-04-13

    Recreational Vehicles LP Gas System Fire Hazard Recall

    Forest River is recalling certain 2021–2023 recreational vehicles because quick disconnect fittings in the LP gas system may crack and leak, creating a fire risk in the presence of an ignition source.

    Product
    COACHMEN — COACHMEN, FOREST RIVER, PRIME TIME FREELANDER, IMPRESSION, LACROSSE, SANIBEL
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0869-2022·2022-04-13

    X-ray System Radiation Dose May Exceed Safety Limits in Some Modes

    Shimadzu SonialVision safire X-ray systems may emit radiation exceeding federal limits during some fluoroscopic procedures. The manufacturer is addressing installation criteria to prevent potential dose violations.

    Product
    MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
    Category
    Medical Device
    Distribution
    37 states
  • HighFDA (Devices)·Z-0863-2022·2022-04-13

    Stradis Medical Anesthesia Kit Recalled Due to Sterility Breach Risk

    Stradis Medical is recalling Lum Lam Custom Pack Anesthesia Kits because the outer bag may be perforated or torn, potentially compromising the kit's sterility. Two lot numbers are affected.

    Product
    Lum Lam Custom Pack - Anesthesia Kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0877-2022·2022-04-13

    Panther Fusion GBS Assay Cartridges Recalled for False Negative Results

    Hologic is recalling 2,061 Panther Fusion GBS Assay Cartridges because samples at or near the limit of detection may produce false negative results, risking missed diagnosis of group B streptococcus infection.

    Product
    Panther Fusion GBS Assay Cartridges 96 Tests, Catalog Number: PRD-04484. in vitro diagnostic test for group b streptococcus.
    Category
    Medical Device
    Distribution
    21 states
  • HighFDA (Devices)·Z-0878-2022·2022-04-13

    DeRoyal ST Neonatal Dressing Change Kit recalled for unvalidated manufacturing test methods

    DeRoyal Industries is recalling 400 DeRoyal ST Neonatal Dressing Change Kits due to uncertain validation of manufacturing test methods. The recall affects kits distributed to healthcare facilities in Florida, New Jersey, and Tennessee.

    Product
    DeRoyal ST Neonatal Dressing Change Kit, containing Prevantics antiseptic (chlorhexidine gluconate and isopropyl alcohol) swab, for preparation of the patient skin prior to surgery or injection.
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-0890-2022·2022-04-13

    Customized Brius orthodontic devices recalled for design control failure

    Brius Technologies is recalling 13 customized metal orthodontic appliances that were sold and distributed prior to completing the required design control process and documentation.

    Product
    Customized Brius Appliances (custom metal orthodontic devices); BRIUS , RX Only, Non-Sterile, Custom Made Device, ID: 52345400UL
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0867-2022·2022-04-13

    X-ray fluoroscopy system may deliver excessive radiation dose

    Shimadzu FLUOROspeed X1 fluoroscopy systems may deliver radiation doses exceeding federal regulatory limits in specific fluoroscopic operating modes due to inadequate installation adjustment criteria.

    Product
    MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
    Category
    Medical Device
    Distribution
    37 states
  • HighFDA (Drugs)·D-0755-2022·2022-04-13

    Pfizer Accuretic Blood Pressure Tablets Recalled for Chemical Impurity

    Pfizer is recalling Accuretic (quinapril HCl/hydrochlorothiazide) tablets nationwide due to unacceptable levels of N-nitroso chemical impurities from manufacturing deviations. The affected lot includes 160 bottles.

    Product
    Accuretic (quinapril HCl/hydrochlorothiazide) Tablets 10 mg/12.5 mg 90 tablets bottle, Rx Only, Distributed by: Parke-Davis, Division of Pfizer Inc, NY, NY 10017, Made in Germany, NDC 0071-0222-23
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0891-2022·2022-04-13

    Abbott ARCHITECT STAT myoglobin calibrators recalled for calibration failure

    Abbott is recalling 402 kits of ARCHITECT STAT myoglobin calibrator that failed stability testing and may produce out-of-specification calibration results. The affected calibrators could impact the accuracy of myoglobin measurements used in clinical diagnostics.

    Product
    ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0741-2022·2022-04-13

    Tetracaine Injection Recalled for Sterility Assurance Manufacturing Defect

    Tetracaine 1% Injectable is being recalled due to sterility assurance failures caused by manufacturing deviations that may compromise product sterility.

    Product
    TETRACAINE 1% Tetracaine HCI Injection, USP, 20mg/2mL (10mg/mL), 10 x 2ml Single Use Vials per box, Rx only, Manufactured for Cameron Pharmaceuticals, LLC., NDC 42494-437-10.
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0898-2022·2022-04-13

    Medline Excelsior Medical Sterile Saline Flush Syringes Recalled for Sterility Failure

    Medline Industries is recalling 7,800 Excelsior Medical 10mL Sterile Field Saline Flush syringes due to sterility failure of the exterior syringe surface, which may prevent the product from being safely used on a sterile field.

    Product
    Excelsior Medical 10mL Sterile Field Saline Flush ZR, 0.9% Sodium Chloride Injection, USP, REF EMZ10091280, Sterile, IV Flush Only, Contents of Package may be dropped on a Sterile Field, 1 syringe/pouch, 35 pouches/box, 8 boxes/case.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0864-2022·2022-04-13

    Total Knee Surgical Kit Recall Due to Potential Sterility Breach

    Stradis Medical is recalling 76 units of Total Knee Surgical Kits with potentially perforated or torn outer bags that may compromise sterility. Distribution occurred in Indiana, Texas, and Pennsylvania.

    Product
    Total Knee Pack - Surgical Kit
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0961-2022·2022-04-13

    Asian Style Noodle Salad recalled for possible plastic contamination

    GH Foods GA is recalling 1,421 units of Asian Style Noodle Salad because an ingredient may contain clear plastic objects. The product was distributed in Georgia and South Carolina.

    Product
    Asian Style Noodle Salad, Net Wt 1lb (16oz) 453 g PERISHABLE, KEEP REFRIGERATED CONTAINS: SOY, TREE NUTS (ALMONDS), WHEAT. Distributed By: Renaissance Food Group, LLC 4222 Moreland Avenue, Conley, GA 30286 UPC 8 26766 14888 1
    Category
    Food
    Distribution
    2 states
  • HighNHTSA·22V245000·2022-04-13

    2021-2022 Volkswagen Atlas Fuel Injector Fire Risk Recall

    Volkswagen is recalling 2021-2022 Atlas and Atlas Cross Sport vehicles because some fuel injectors lack filters, potentially causing fuel leaks and increasing fire risk. Dealers will replace affected injectors at no cost.

    Product
    VOLKSWAGEN — VOLKSWAGEN ATLAS CROSS SPORT, ATLAS
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0969-2022·2022-04-13

    Thai Noodle Salad Kit Recalled for Potential Plexiglass Contamination

    Taylor Farms Southwest is recalling Thai Noodle Salad Kits (108 oz bags) due to potential contamination with plexiglass pieces supplied by an ingredient vendor. Consumers should not consume the affected product.

    Product
    Thai Noodle Salad Kit 108 oz. bag (behind the counter salad), UPC 0 30223 11270 9, Taylor Farms Southwest, 859 S. 86th Ave, Tolleson, AZ 85353. Perishable, refrigerate until consumption.
    Category
    Food
    Distribution
    1 state
  • HighFDA (Devices)·Z-0861-2022·2022-04-13

    Defibrillator devices unapproved for US market recalled by FDA

    Remote Diagnostic Technologies recalls Tempus LS defibrillators not approved for US distribution. Two units were distributed in Illinois and Kansas.

    Product
    Tempus LS, Type DGE1, CAT 00-3010, REF 3.940590, Software version 1.3.5, UDI: (01)07613365001693
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0742-2022·2022-04-13

    Drug recall: Papaverine hydrochloride injection due to sterility concerns

    Vitae Enim Vitae Scientific, Inc. is recalling papaverine hydrochloride injection nationwide due to manufacturing deviations that compromise sterility assurance in this intravenous drug product.

    Product
    PAPAVERINE HYDROCHLORIDE — PAPAVERINE HYDROCHLORIDE (PAPAVERINE HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0868-2022·2022-04-13

    X-ray System Models Recalled for Potential Radiation Dose Exceeding Standards

    Shimadzu Medical Systems recalls 176 SonialVision G4 X-ray systems that may emit radiation doses exceeding federal standards in some fluoroscopic mode cases. No injuries have been reported.

    Product
    MODEL: X-RAY TV SYSTEM SONIALVISION G4
    Category
    Medical Device
    Distribution
    37 states
  • HighFDA (Devices)·Z-0874-2022·2022-04-13

    Philips Allura and Azurion X-ray systems may display incorrect previous images

    Philips Allura and Azurion X-ray systems running StentBoost Live R2.0 may display images from previous scans or other patients due to a software defect. 338 units are affected and globally distributed.

    Product
    Philips Allura (Model Numbers 722010, 722012, 722026, 722027, 722028, 722029, 722035) and Philips Azurion (Model Numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228) running StentBoost Live (Model Number…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0859-2022·2022-04-13

    GE SIGNA Excite 3T MRI Systems Recalled Over Potential Image Flip

    GE Healthcare is recalling 55 SIGNA Excite 3T MRI systems due to potential for images to be flipped left to right, which could affect diagnostic accuracy. The affected systems were distributed worldwide.

    Product
    SIGNA Excite 3T. Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    Distributed nationwide