The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means the agency’s own highest classification or reported deaths or serious injuries, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

3051–3075 of 90254

  • HighFDA (Devices)·Z-1654-2026·2026-04-08

    Paradise renal denervation catheter unit distributed despite scrapping designation

    One Paradise Ultrasound Renal Denervation catheter was distributed although it was a nonconforming product that should have been scrapped and not used in humans.

    Product
    Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1746-2026·2026-04-08

    MiniMed 740G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 740G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1692-2026·2026-04-08

    Vue Motion V12 may display cine image frames out of sequence

    Philips is recalling Vue Motion V12 because mis-ordered frames during dynamic cine runs may cause image frames to display out of sequence.

    Product
    Vue Motion V12. Product Number: 1017979.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1696-2026·2026-04-08

    da Vinci monopolar curved scissors recalled over frayed grip cables

    Intuitive Surgical is recalling da Vinci S and Si Monopolar Curved Scissors Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1656-2026·2026-04-08

    RefleXion X1 may deliver radiotherapy dose to the incorrect location

    RefleXion Medical is recalling the X1 Radiotherapy System because a software defect can mean angular roll corrections are not carried through to treatment delivery.

    Product
    RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
    Category
    Medical Device
    Distribution
    8 states
  • HighFDA (Food)·H-0646-2026·2026-04-08

    Coconut Drink 1 recalled due to undeclared milk allergen

    HEYTEA USA INC is recalling Coconut Drink 1 (1L cartons) because the product contains undeclared milk. Consumers with milk allergies or sensitivities are at risk.

    Product
    Coconut Drink 1; NET VOLUME: 1L; Ingredients: Coconut water, freshly squeezed coconut gravy, white granulated sugar, emulsion stabilizer [sodium caseinate, mono- and diglycerides fatty acid esters, sucrose fatty acid esters]; Storage Condition: Room temperature; Manufactured in…
    Category
    Food
    Distribution
    8 states
  • HighFDA (Devices)·Z-1743-2026·2026-04-08

    MiniMed 640G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 640G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1667-2026·2026-04-08

    Philips Allura Xper FD10 OR Table X-ray may not start reliably

    Philips is recalling the Allura Xper FD10 OR Table because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1693-2026·2026-04-08

    Sterile radiology procedure kits may have a breached sterile pouch

    One lot of Sterile Radiology Procedure Kits, model DYNDH1491B, is being recalled because of a potential breach in the sterile pouch packaging that may compromise sterility.

    Product
    Sterile Radiology Procedure Kits, Model Number DYNDH1491B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1767-2026·2026-04-08

    GE SIGNA Premier MRI magnets may have ferrous fittings fitted

    SIGNA Premier MRI systems are being recalled because they may have been built or serviced with ferrous steel fittings instead of non-ferrous brass on the magnet rear.

    Product
    SIGNA Premier systems
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1738-2026·2026-04-08

    Civco eTRAX 18G needle sensor may misreport needle tip position

    Civco is recalling the eTRAX Needle Sensor 18G for Aurora trackers because an inspection and programming error may cause the needle tip position to be incorrectly identified.

    Product
    eTRAX Needle Sensor - 18G (for Aurora Trackers), Part Number 667-159
    Category
    Medical Device
    Distribution
    3 states
  • HighFDA (Devices)·Z-1671-2026·2026-04-08

    Philips Allura Xper FD20/15 X-ray may not start from foot switch

    Philips is recalling the Allura Xper FD20/15 because X-ray imaging may not be initiated, or may occur intermittently, when using the wired foot switch.

    Product
    Allura Xper FD20/15; System Code: 722058;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1690-2026·2026-04-08

    Medline ENFit G-tube connectors may leak from dimensional specification failure

    Medline is recalling ENFit G-tube connectors because they were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices.

    Product
    ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Drugs)·D-0420-2026·2026-04-08

    Prazosin hydrochloride tablets recalled over nitrosamine impurity above limits

    Five lots of Prazosin Hydrochloride tablets are being recalled after detection of the nitrosamine impurity N-nitroso-prazosin impurity C above acceptable limits.

    Product
    PRAZOSIN HYDROCHLORIDE — PRAZOSIN HYDROCHLORIDE (PRAZOSIN HYDROCHLORIDE)
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1750-2026·2026-04-08

    Paradigm REAL-Time Revel pump delivery varies with pump height changes

    Medtronic is recalling Paradigm REAL-Time Revel insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0637-2026·2026-04-08

    Divided Sunset Multi Collagen Peptides Missing Allergen Labeling

    TG FOODS INC is recalling Divided Sunset Multi Collagen Peptides (8 oz) because the product contains eggshell membrane collagen and wild-caught marine collagen but fails to declare EGG and the specific fish species on the label.

    Product
    Divided Sunset Multi Collagen Peptides, 8 Ounce, Stand-up Pouch, UPC Code: 8 50005 60689 5
    Category
    Food
    Distribution
    28 states
  • HighFDA (Food)·H-0660-2026·2026-04-08

    Imu-Tek Colostrum-5 Capsules Recalled for Potential Under-Processing

    Imu-Tek Animal Health is recalling Imu-Tek Colostrum-5 120 Capsules (Lot #013, Expiration 092826) because the product is potentially under-processed. The product was distributed across 29 states.

    Product
    Imu-Tek Colostrum-5 120 Capsules; 30% IgG, 2 capsules 2x/day, orally with 8 ounces of water, support a healthy immune system, 3 year shelf life, plastic bottle, inner liner seal and shrink band seal, 500 mg capsule, 7 38654 00026 7, Imu-Tek, Inc. Ft. Collins, CO 80524.
    Category
    Food
    Distribution
    30 states
  • HighFDA (Devices)·Z-1683-2026·2026-04-08

    CPX 4 medium height tissue expander infusion needles may be blunt

    Mentor is recalling CPX 4 US Breast Tissue Expanders with Suture Tabs in the smooth medium height range because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800cc
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1740-2026·2026-04-08

    MiniMed 630G insulin pump delivery varies with pump height changes

    Medtronic is recalling MiniMed 630G insulin pumps because raising or lowering the pump relative to the infusion site can cause unintended over- or under-delivery of insulin.

    Product
    MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1701-2026·2026-04-08

    da Vinci permanent cautery hook recalled over frayed pitch cables

    Intuitive Surgical is recalling da Vinci S and Si Permanent Cautery Hook Instruments following increased complaints regarding frayed or broken pitch cables on reusable surgical instruments.

    Product
    Brand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S, Si Permanent Cautery Hook Instrument Model/Catalog Number: 420183 Software Version: NA Product Description: EndoWrist Instruments or da Vinci S/Si Reusable instruments, including blunt and…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1698-2026·2026-04-08

    da Vinci mega needle driver recalled over broken or frayed cables

    Intuitive Surgical is recalling da Vinci S and Si Mega Needle Driver Instruments following increased complaints of broken or frayed grip cables in reusable instruments.

    Product
    Brand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Model/Catalog Number: 420194 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1710-2026·2026-04-08

    Medline bronchoscopy kits contain Olympus biopsy valves that shed fragments

    Medline is recalling six bronchoscopy and thoracic kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.

    Product
    Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6.…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1766-2026·2026-04-08

    Oxoid Microbact Listeria kit may identify the incorrect organism

    Oxoid, part of Thermo Fisher Scientific, is recalling one lot of the Microbact 12L Listeria identification kit because it may identify the incorrect organism.

    Product
    OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1684-2026·2026-04-08

    CPX 4 tall height tissue expander infusion needles may be blunt

    Mentor is recalling CPX 4 US Breast Tissue Expanders with Suture Tabs in the smooth tall height range because the infusion sets may contain a dull or blunt needle tip.

    Product
    CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-113TH, SCPX-123TH, SCPX-140TH, SCPX-146TH Smooth Tall Height Tissue Expander, 350cc, 450cc, 650cc, 750cc
    Category
    Medical Device
    Distribution
    Distributed nationwide