Medline bronchoscopy kits contain Olympus biopsy valves that shed fragments
Medline is recalling six bronchoscopy and thoracic kits because they contain recalled Olympus biopsy valves, where rubber fragments may detach from the valve slit during use.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported: fragments may lodge in a patient's airway and require removal, and no injuries are stated in the source.
Plain-English summary
Medline Industries, LP is recalling kits containing Olympus biopsy valves, covering the Bronch Procedure Kit (DYKE2091), Bronchoscopy kits (DYNJ900898G, DYNJ901922I), Bronchoscopy Pack (DYNJ38313B), Bronchoscopy Pack 0120367-LF (DYNJ32750G, DYNJ32750I), WR Thoracic Kit (DYNJ911252) and a Laryngoscopy/Bronchoscopy kit. About 14,379 kits were distributed across thirteen US states. The FDA has classified this recall as Class II.
The kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences of fragment detachment include a foreign body in the patient's tracheobronchial tree, which may require intervention for removal, along with inflammatory response, hypoxia and prolongation of the procedure.
Each kit carries its own UDI-DI and lot numbers, listed in the FDA enforcement record.
Facilities can identify affected kits by the SKU, UDI and lot numbers, and may contact Medline Industries, LP for further information.
The recalled product
- Product
- Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRON
- Manufacturer
- Medline Industries, LP
- Category
- Medical Device — Endoscopy
- Affected units
- 14,379
Distribution
Part of a larger recall action
This product is one of 7 recalled by Medline Industries, LP under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 7 products in this recall →Related recalls
Same manufacturer · Medline Industries, LP
See all →- HighMedline Recalls Cataract Preop Kits with Potentially Contaminated Applicators
FDA (Devices) · 2026-09-09
- SevereMedline C-Section Prep Kits Recalled for Potential Sterility Breach
FDA (Devices) · 2026-09-09
- HighMedline Blood Culture Kits Recalled Due to Potential Seal Defects
FDA (Devices) · 2026-09-09
- SevereMedline Eye and Cataract Kits Recalled for Potential Sterility Breach
FDA (Devices) · 2026-09-09
- SevereMedline Ultrasound Breast Pack Recalled for Potential Sterility Breach
FDA (Devices) · 2026-09-09
Same hazard · fragment detachment
See all →- HighMedline laryngoscopy bronchoscopy kits contain Olympus valves shedding rubber fragments
FDA (Devices) · 2026-04-08
- HighMedline anterior hip pack contains recalled Olympus biopsy valves
FDA (Devices) · 2026-04-08
- HighAirway exam kits contain recalled Olympus biopsy valves shedding fragments
FDA (Devices) · 2026-04-08
- HighMedline endoscopy kits contain recalled Olympus biopsy valves shedding fragments
FDA (Devices) · 2026-04-08
- HighMedline bronchoscopy packs contain recalled Olympus biopsy valves shedding fragments
FDA (Devices) · 2026-04-08