MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
Pending LLM rewrite. Source: FDA_DEVICE Z-1746-2026.
Severity not yet assessed
We have not scored this recall against our rubric yet, so we are not showing a severity rating or our usual guidance. The agency’s own notice below is the authoritative source — read it first.
Plain-English summary
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
The recalled product
- Product
- MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
- Manufacturer
- Medtronic MiniMed, Inc.
- Category
- Medical Device — Devices
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 763000504915
- 763000505073
- 763000484637
- 763000505042
- 763000504861
- 763000629984
- 763000504946
- 763000504878
- 763000559557
- 763000504960
- 763000505080
- 763000505066
- 763000629991
- 763000505004
- 763000504922
- 763000504892
- 763000505059
- 763000504977
- 763000505028
- 763000505172
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- HighMedtronic DLP One-Piece Pediatric Arterial Cannula Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- SevereAbiomed Impella CP Smart Assist Set Heart Pump Introducer Sheath Leakage
FDA (Devices) · 2026-07-22
- SevereMedtronic Bio-Medicus Adult Cannula Kit Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22