Zimmer Biomet Recalls Orthopedic Drill Bits Due to Fracture Risk
Zimmer Biomet is recalling 240 ECT Internal Fracture Fixation Quick-Connect Drills because raw material anomalies may cause the instruments to fracture during use.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (surgical instrument fracture) where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Zimmer Biomet, Inc. is recalling 240 units of the ECT Internal Fracture Fixation Quick-Connect Drill 2.7 mm Diameter 10 cm Length (Item Number/EDI 00241002700). These nonsterile orthopedic surgical instruments were distributed nationwide in the United States and internationally to Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan, and Trinidad.
The recall was initiated because various trauma and extremity instruments, including these drill bits, were manufactured from raw material with an anomaly. This anomaly may be on or below the surface and can increase the risk of instrument fracture. The affected lot numbers are 63306444, 63315170, 63350728, 63364047, 63377460, 63474111, 63474113, 63499659, and 63499667.
Healthcare providers and facilities should check their inventory for these specific item numbers and lot numbers. If the recalled products are found, they should be removed from service and returned to Zimmer Biomet, Inc. according to the company's instructions.
The recalled product
- Product
- ECT Internal Fracture Fixation Quick-Connect Drill 2.7 mm Diameter 10 cm Length, Item Number/EDI 00241002700, Nonsterile. orthopedic surgical instrument.
- Manufacturer
- Zimmer Biomet, Inc.
- Hazard
- Affected units
- 240
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (8)
- Lot Numbers: 63306444
- 63315170
- 63350728
- 63364047
- 63377460
- 63474111
- 63474113
- 63499659 & 63499667.
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 60 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 60 products in this recall →Related recalls
Same manufacturer · Zimmer Biomet, Inc.
See all →- HighZimmer Biomet Recalls Spinal Rod Cutters Due to Fracture Risk
FDA (Devices) · 2020-12-30
- ModerateZimmer Biomet Recalls Polaris Spinal System Screws for Labeling Update
FDA (Devices) · 2020-12-23
- HighZimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Concerns
FDA (Devices) · 2020-11-11
Same hazard · instrument fracture
See all →- HighZimmer Biomet Recalls 1.0mm Drill Bits Due to Fracture Risk
FDA (Devices) · 2018-02-14
- HighZimmer Biomet Recalls Herbert Cannulated Bone Screw System Drills
FDA (Devices) · 2017-10-04
- HighZimmer Biomet Recalls Orthopedic Drill Bits Due to Fracture Risk
FDA (Devices) · 2017-10-04
- SevereZimmer Biomet Recalls Orthopedic Drill Bits Due to Fracture Risk
FDA (Devices) · 2017-10-04
- HighZimmer Biomet Recalls Orthopedic Drill Bits Due to Fracture Risk
FDA (Devices) · 2017-10-04
Similar recalls
- HighZimmer Biomet Recalls 1.0mm Drill Bits Due to Fracture Risk
FDA (Devices) · 2018-02-14
- HighZimmer Biomet Recalls Spinal Rod Cutters Due to Fracture Risk
FDA (Devices) · 2020-12-30
- ModerateZimmer Biomet Recalls Affixus Hip Fracture Nails Due to Assembly Defect
FDA (Devices) · 2018-02-07
- SevereZimmer Biomet Recalls Antegrade Femoral Nail Instrument Kits
FDA (Devices) · 2018-03-07
- HighZimmer Biomet Recalls Connecting Bolts Due to Material Error
FDA (Devices) · 2018-08-29
- HighZimmer Biomet Recalls AFFIXUS Hip Fracture Nail Due to Sizing Defect
FDA (Devices) · 2017-06-07