Zimmer Biomet Recalls Affixus Hip Fracture Nails Due to Set Screw Defect
Zimmer Biomet is recalling Affixus Hip Fracture Nails because the set screw cannot advance or reverse from its original position.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the defect in a structural orthopedic implant represents a risk of harm where injury has not yet been reported.
Plain-English summary
Zimmer Biomet, Inc. is recalling 5,947 units of Affixus Hip Fracture Nail Left 130° 11 mm x 460 mm (Item Number: 814611460). The recall is due to a defect where the set screw is not able to advance or reverse from the original position in the nail.
The affected products were distributed to 38 U.S. states and Puerto Rico, as well as Chile, Ecuador, Japan, and the Netherlands. The specific lot codes associated with this recall are 171830 and 712290.
Healthcare providers and patients should contact Zimmer Biomet for further instructions regarding the return or replacement of these devices.
The recalled product
- Product
- Affixus¿ Hip Fracture Nail Left 130¿ 11 mm x 460 mm, Item Number: 814611460
- Manufacturer
- Zimmer Biomet, Inc.
- Hazard
- Affected units
- 5,947
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- 171830 712290
Distribution
Part of a larger recall action
This product is one of 129 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 129 products in this recall →Related recalls
Same manufacturer · Zimmer Biomet, Inc.
See all →- HighZimmer Biomet Recalls Spinal Rod Cutters Due to Fracture Risk
FDA (Devices) · 2020-12-30
- ModerateZimmer Biomet Recalls Polaris Spinal System Screws for Labeling Update
FDA (Devices) · 2020-12-23
- HighZimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Concerns
FDA (Devices) · 2020-11-11
Same hazard · mechanical failure
See all →- SevereCanon Medical Alphenix INFX-8000C/B5 Recalled for Tabletop Shift Risk
FDA (Devices) · 2026-10-14
- ModerateLeMaitre TufTex Over-the-Wire Embolectomy Catheter Recalled for Balloon Failure
FDA (Devices) · 2026-10-14
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
Similar recalls
- HighZimmer Biomet Recalls Affixus Hip Fracture Nails Due to Set Screw Defect
FDA (Devices) · 2018-11-21
- HighZimmer Biomet Recalls Affixus Hip Fracture Nails Due to Set Screw Issue
FDA (Devices) · 2018-11-21
- HighZimmer Biomet Recalls Affixus Hip Fracture Nail Due to Set Screw Defect
FDA (Devices) · 2018-11-21
- HighZimmer Biomet Recalls Affixus Hip Fracture Nail Due to Set Screw Issue
FDA (Devices) · 2018-11-21
- HighZimmer Biomet Recalls Affixus Hip Fracture Nails Due to Set Screw Malfunction
FDA (Devices) · 2018-11-21
- HighZimmer Biomet Recalls Affixus Hip Fracture Nail Due to Assembly Defect
FDA (Devices) · 2018-02-07