Siemens Dimension Vista CHEM 1 CAL Recall for Magnesium Calibration Error
Siemens is recalling Dimension Vista CHEM 1 CAL lot 3GM081 because it causes under-recovery of Magnesium in patient and QC samples.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any hospitalizations, injuries, or deaths, fitting the criteria for a moderate severity recall involving a diagnostic accuracy issue without reported harm.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling the Dimension Vista CHEM 1 CAL, lot 3GM081. This in vitro diagnostic product is used for the calibration of various methods, including Magnesium (MG), on the Dimension Vista System. The recall involves 9,769 total units, with 6,000 distributed in the United States and the remainder distributed internationally to countries including Australia, Austria, Belgium, Canada, and others.
The recall was initiated after Siemens received complaints regarding low recovery of Magnesium in Quality Control (QC) and patient samples following calibration with the affected lot. An investigation by Siemens confirmed that the product causes QC and patient samples to under-recover Magnesium by 0.25 mg/dL [0.10 mmol/L] across the assay range.
Healthcare facilities and laboratories that have used or possess this specific lot of the calibration product should stop using it and contact Siemens Healthcare Diagnostics for instructions on return or replacement. The affected lot has an expiration date of 07-01-2014.
The recalled product
- Product
- Siemens Dimension Vista¿ CHEM 1 CAL, lot 3GM081. The CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyroxine Uptake (TU), Blood Urea…
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- lot 3GM081
- exp. 07-01-2014
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · diagnostic accuracy
See all →- HighEmano Flow Uroflowmetry App Recalled Over Biased Flow Estimates
FDA (Devices) · 2026-07-29
- HighYeastone Broth recalled because it may report incorrect susceptibility results
FDA (Devices) · 2026-04-22
- HighPreeclampsia growth factor assay recalled over out-of-range quality control values
FDA (Devices) · 2026-04-15
- HighMicroScan multidrug resistance test panels recalled over incorrect processing
FDA (Devices) · 2026-04-15
- HighMicroScan susceptibility panels recalled over missing drug and doubled antibiotic
FDA (Devices) · 2026-04-15
Similar recalls
- HighSiemens Dimension Vista Chemistry 1 Calibrator Recall for Low Calcium Recovery
FDA (Devices) · 2016-11-30
- ModerateSiemens Dimension Vista Enzyme 1 Calibrator Recalled for QC Shifts
FDA (Devices) · 2014-03-05
- HighSiemens Dimension Vista Calcium Flex Reagent Cartridge Recalled for Low Results
FDA (Devices) · 2018-05-09
- ModerateSiemens Dimension Vista Protein 1 Calibrator Recalled for Inaccurate Ceruloplasmin Results
FDA (Devices) · 2017-12-06
- HighSiemens Dimension Vista Magnesium Reagent Cartridge Recall for Low Results
FDA (Devices) · 2015-12-02
- HighSiemens Magnesium Flex reagent cartridge recalled for imprecision and assay errors
FDA (Devices) · 2022-04-06