Siemens Dimension Vista Magnesium Reagent Cartridge Recall for Low Results
Siemens Healthcare Diagnostics is recalling 3,601 Dimension Vista magnesium reagent cartridges due to erroneous low results that may affect patient and quality control testing.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of harm (erroneous low magnesium results) where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 3,601 units of the Dimension Vista System (MG) Flex reagent cartridge, Catalog number K3057, SMN 10445158, Lot Number 15063BA. The product is used for the quantitative measurement of magnesium in human serum, plasma, and urine on the Dimension Vista System.
The recall was initiated because of erroneous low results on a small number of reagent wells. This issue creates the potential for under-recovery in both quality control (QC) and patient sample results. The under-recovery ranged from -0.3 mg/dL [0.12 mmol/L] to -1.6 mg/dL [0.66 mmol/L], which is equivalent to a maximum negative bias of -88%.
The affected products were distributed nationwide in the United States, including DC and specific states, as well as in Canada. No specific instructions for consumers are provided in the source text, but the recall is managed by the recalling firm, Siemens Healthcare Diagnostics, Inc.
The recalled product
- Product
- Dimension Vista System (MG) ; Flex reagent cartridge Catalog number K3057, SMN 10445158. For the quantitative measurement of magnesium in human serum, plasma and urine on the Dimension Vista System.
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 3,601
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Lot Number 15063BA
Distribution
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · erroneous results
See all →- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighChemistry analyzer control modules recalled over software that may give erroneous results
FDA (Devices) · 2025-12-24
- HighLaboratory automation recalled over unlabelled or mismatched secondary sample tubes
FDA (Devices) · 2025-12-03
- HighBlood lead test kits recalled over mislabeled calibration buttons
FDA (Devices) · 2025-12-03
- HighLaboratory automation recalled over sample tubes with mismatched labels
FDA (Devices) · 2025-12-03
Similar recalls
- HighSiemens Magnesium Flex reagent cartridge recalled for imprecision and assay errors
FDA (Devices) · 2022-04-06
- HighSiemens Dimension Vista Calcium Flex Reagent Cartridge Recalled for Low Results
FDA (Devices) · 2018-05-09
- HighSiemens Dimension Vista Calcium Reagent Cartridge Recall for Erroneous Results
FDA (Devices) · 2016-10-12
- HighSiemens Magnesium Test Cartridges May Produce Erroneous Results Without Alert
FDA (Devices) · 2022-10-05
- HighSiemens magnesium reagent cartridges recalled for false measurement results
FDA (Devices) · 2025-01-01
- HighSiemens Recalls Dimension Vista B2-Microglobulin Reagent Cartridges
FDA (Devices) · 2015-09-02