Siemens syngo.via RT Image Suite Software Recall for Geometry Errors
Siemens is recalling syngo.via RT Image Suite software versions VB30 and VB40 because modified MR image orientations may result in incorrect geometry and radiation treatment plans.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (radiation treatment planning) where injury has not yet been reported, and the source explicitly states the issue has never been reported, fitting the 'High' severity criteria for theoretical hazards without confirmed illness.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 21 units of the syngo.via RT Image Suite software, specifically versions VB30 and VB40. The recall was initiated because the software may fail to recognize adapted acquisition planes when users modify the original image orientation of standard MR protocols for Synthetic CT processing. This defect can result in images with incorrect geometry.
If these distorted images are exported to a treatment planning system, there is a risk of an incorrectly calculated radiation treatment plan. The source text notes that the occurrence of this issue is very unlikely and has never been reported so far. The affected units were distributed nationwide in the United States, including in the states of AZ, MA, WI, FL, PA, TX, MD, LA, MN, MI, NY, SC, OH, GA, CA, and IL.
Healthcare facilities using this software should contact Siemens Medical Solutions USA, Inc. for instructions on how to address the software defect. The recall is classified as Class II by the FDA.
The recalled product
- Product
- syngo.via RT Image Suite with software versions syngo.via VB30 or VB40
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Category
- Medical Device — Software
- Affected units
- 21
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- syngo.via VB30 or VB40: All Lots Model #: 10496180
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · incorrect treatment plan
See all →- HighRaySearch Recalls RayStation and RayPlan Software Due to Treatment Plan Errors
FDA (Devices) · 2020-12-23
- HighRayStation software recall for potential incorrect treatment plan references
FDA (Devices) · 2020-12-02
- HighStryker Recalls Hoffmann LRF Hexapod Software Due to Calculation Error
FDA (Devices) · 2018-10-03
- HighNucletron Oncentra Brachy 4.5 Software Recalled for Incorrect Source Step Size
FDA (Devices) · 2017-11-01
Similar recalls
- SevereSiemens Syngo.via RT Image Suite Software Recall for Calibration Offset Error
FDA (Devices) · 2020-09-30
- HighRadiation Therapy Software Risk: Data Loss in Treatment Planning
FDA (Devices) · 2021-09-22
- HighRadiation therapy software update displays incorrect laser positioning
FDA (Devices) · 2022-05-04
- ModerateSiemens Syngo Imaging Software Recall for Version VB36D_HF02
FDA (Devices) · 2015-12-02
- ModerateSiemens syngo.plaza Radiological Image Processing System Software Update Recall
FDA (Devices) · 2012-12-12
- HighSiemens Medical Solutions Recalls syngo.via Cardiac Planning Software
FDA (Devices) · 2019-04-17