Allura CV20 systems recalled over internal component deterioration disabling motorized movement
Philips Allura CV20 systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 46 units are affected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported; loss of motorized movement during an interventional procedure is a potential harm and no injuries are reported.
Plain-English summary
The Philips Allura CV20, product code 722031, is being recalled. Affected serial numbers include 155, 150, 87, 88, 109, 97, 138, 86, 152, 135, 153, 35 and 76, with the full list in the recall notice. The recall covers 46 units, all outside the United States, within a worldwide distribution.
Systems may not perform as intended because of deterioration of certain internal components: the CMOS battery, hard disk drive and/or power supply unit. If this deterioration occurs, motorized movements of the system become unavailable, leaving only manual movements of the stand and table. The FDA classified the action as Class II.
Imaging sites operating these systems should follow the manufacturer's instructions for inspection and correction.
The recalled product
- Product
- Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Category
- Medical Device — Imaging System
- Affected units
- 46
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Product Code: 722031
- UDI-DI: N/A
- Serial numbers: 155
- 150
- 87
- 88
- 109
- 97
- 138
- 86
- 152
- 135
- 153
- 35
- 76
- 133
- 173
- 124
- 92
- 166
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 13 recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 13 products in this recall →Related recalls
Same manufacturer · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
See all →- HighPhilips Azurion Rystem X-Ray System Software Recall for Image Quality Issues
FDA (Devices) · 2026-08-19
- HighPhilips Allura Xper FD Fluoroscopic X-Ray System Recalled for Firmware Issue
FDA (Devices) · 2026-08-19
- HighPhilips Azurion Fluoroscopy System Recall for Software Image Quality Issues
FDA (Devices) · 2026-08-19
- HighPhilips Fluoroscopy System Recall Over Hard Disk Drive Issue
FDA (Devices) · 2026-08-12
- SeverePhilips Allura X-Ray Systems Recalled Over Unauthorized Hard Disk Drives
FDA (Devices) · 2026-08-12
Same hazard · component deterioration
See all →- HighAllura Xper FD10 OR Table recalled over ageing power supply and drive
FDA (Devices) · 2025-10-01
- HighAllura Xper FD20 systems recalled over component deterioration halting motorized movements
FDA (Devices) · 2025-10-01
- HighAllura Xper FD20/20 OR Table recalled over deteriorating internal system components
FDA (Devices) · 2025-10-01
- HighAllura Xper FD20/20 systems recalled over battery, drive and power supply deterioration
FDA (Devices) · 2025-10-01
- ModerateAllura Xper FD20/15 X-ray System Motorized Movement Component Deterioration
FDA (Devices) · 2025-10-01
Similar recalls
- ModeratePhilips Allura CV20 X-ray System Recalled for Capacitor Failure
FDA (Devices) · 2020-03-25
- ModerateAllura Xper CV20 X-ray systems recalled after BIOS battery halts startup
FDA (Devices) · 2025-10-08
- SeverePhilips Allura CV20 Patient Table: Fall Risk from Mattress Defect
FDA (Devices) · 2025-02-26
- ModerateAllura Xper FD20 X-ray systems recalled over premature BIOS battery depletion
FDA (Devices) · 2025-10-08
- ModerateAllura Xper FD20/20 X-ray systems recalled for BIOS battery startup failure
FDA (Devices) · 2025-10-08
- HighPhilips Allura Medical Imaging Systems Recalled for Image Interpretation Error
FDA (Devices) · 2016-02-10