The Recall Desk

Hazard

Interventional Imaging recalls

11 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
11
Critical
0
Severe
0
Most recent
2025-10-01

Of the 11 interventional imaging recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all interventional imaging recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–11 of 11

  • HighFDA (Devices)·Z-2631-2025·2025-10-01

    Allura Xper FD20/20 OR Table recalled over deteriorating internal system components

    Philips Allura Xper FD20/20 OR Table systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 3 units are affected.

    Product
    Allura Xper FD20/20 OR Table; Product Code: 722039; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2623-2025·2025-10-01

    Allura Xper FD10 systems recalled over ageing internal components stopping motorised movement

    Philips Allura Xper FD10 systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 1,117 units are affected.

    Product
    Allura Xper FD10 Product Codes: (1)722010, (2) 722026, (3) 722003; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2632-2025·2025-10-01

    Allura Xper FD20/10 systems recalled over deterioration disabling motorized movement

    Philips Allura Xper FD20/10 systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 45 units are affected.

    Product
    Allura Xper FD20/10; Product Code: 722029; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2627-2025·2025-10-01

    Allura Xper FD20 Biplane recalled over component deterioration affecting motorized movements

    Philips Allura Xper FD20 Biplane systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 268 units are affected.

    Product
    Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2630-2025·2025-10-01

    Allura Xper FD20/20 systems recalled over battery, drive and power supply deterioration

    Philips Allura Xper FD20/20 systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 75 units are affected.

    Product
    Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2626-2025·2025-10-01

    Allura Xper FD20 systems recalled over component deterioration halting motorized movements

    Philips Allura Xper FD20 systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 2,114 units are affected.

    Product
    Allura Xper FD20 Product Codes: (1)722012, (2) 722028, (3) 722006; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2629-2025·2025-10-01

    Allura Xper FD20 Biplane OR Table recalled over internal component deterioration

    Philips Allura Xper FD20 Biplane OR Table systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 5 units are affected.

    Product
    Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2635-2025·2025-10-01

    Allura CV20 systems recalled over internal component deterioration disabling motorized movement

    Philips Allura CV20 systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 46 units are affected.

    Product
    Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2634-2025·2025-10-01

    Allura Xper FD20/15 OR Table recalled over internal component deterioration

    The Philips Allura Xper FD20/15 OR Table is recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. One unit is affected.

    Product
    Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2628-2025·2025-10-01

    Allura Xper FD20 OR Table recalled over CMOS battery and drive deterioration

    Philips Allura Xper FD20 OR Table systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 174 units are affected.

    Product
    Allura Xper FD20 OR Table Product Codes: (1)722023, (2) 722035, (3) 722015; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2625-2025·2025-10-01

    Allura Xper FD10 OR Table recalled over ageing power supply and drive

    Philips Allura Xper FD10 OR Table systems are recalled because deterioration of the CMOS battery, hard disk drive and/or power supply may make motorized movements unavailable. 2 units are affected.

    Product
    Allura Xper FD10 OR Table Product Codes: (1)722022, (2) 722033; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
    Category
    Medical Device
    Distribution
    Distributed nationwide