Landauer nanoDot radiation dosimeters may be outside accuracy specifications
Landauer is recalling nanoDot D2DNS radiation dosimeters that may be outside the specified +/-5.5% accuracy range. The issue involves a potential non-conformance in the optical measurement material.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II FDA recall. No illnesses or injuries have been reported. The hazard is measurement inaccuracy in radiation dosimeters, categorized as risk-of-harm without reported injury, which the rubric caps at severity level 3.
Plain-English summary
Landauer is recalling its nanoDot D2DNS radiation monitoring dosimeters, Model Numbers 18130-000, 18130-KIT, and 18130-SET. These devices measure radiation exposure and are used with microSTAR readers.
Some devices may be outside the specified accuracy range of +/-5.5%. The issue is a potential non-conformance in the Optical Stimulated Luminescence (OSL) material, which emits fluorescence when exposed to light after receiving radiation exposure. The reader uses this fluorescence signal to calculate the radiation dose received.
Approximately 3,570 units have been distributed worldwide (all batch numbers under UDI/DI 0860003399903). The FDA classified this as a Class II recall. No illnesses or injuries have been reported related to this issue.
The recalled product
- Product
- QC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18130-000; b) 18130-KIT; c) 18130-SET; radiation monitoring dosimeter used with the microSTAR readers
- Manufacturer
- Landauer
- Affected units
- 3,570
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI/DI 0860003399903
- all batch numbers
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 38 recalled by Landauer under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 38 products in this recall →Related recalls
Same manufacturer · Landauer
See all →- ModerateLandauer microStar Reader dosimetry system discontinued for medical applications
FDA (Devices) · 2024-03-27
- HighLandauer microSTARii reader recalled for measurement inaccuracy
FDA (Devices) · 2024-01-03
- HighLandauer nanoDot D2DXS Dosimeter Recall for Measurement Accuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot A Dosimeters Recalled for Potential Measurement Inaccuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot Radiation Dosimeters Recalled Due to Potential Accuracy Issues
FDA (Devices) · 2023-09-20
Same hazard · measurement inaccuracy
See all →- HighCDI OneView System BPM Probe temperature sensor measurement inaccuracy
FDA (Devices) · 2025-07-02
- ModeratePhilips Tempus Pro Patient Monitor Software Update for Operational Issues
FDA (Devices) · 2025-05-14
- SevereVentilator Model LD2110X23B Recalled for Potential Sensor Measurement Failure
FDA (Devices) · 2024-11-20
- SevereAeris EVO Ventilator May Develop Flow Sensor Fouling from Aerosols
FDA (Devices) · 2024-11-20
- SevereTrilogy EV300 ventilator: potential flow sensor malfunction from aerosol accumulation
FDA (Devices) · 2024-11-20