Radiation Dosimeter May Provide Inaccurate Dose Measurements
Landauer is recalling certain nanoDot D2DNS radiation dosimeters because they may measure radiation exposure outside the specified accuracy range. Approximately 5,958 units worldwide are affected.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This Class II medical device may provide inaccurate radiation measurements. Although the potential non-conformance is not confirmed, the risk of measurement inaccuracy in a dosimeter creates a risk-of-harm scenario for workers relying on accurate exposure monitoring.
Plain-English summary
Landauer is recalling Calibrate (80 kVp) nanoDot D2DNS radiation monitoring dosimeters (Model Numbers 04217-000, 04217-KIT, and 04217-SET) used with microSTAR readers. Approximately 5,958 units have been distributed worldwide. The devices are designed to measure occupational radiation exposure.
Some nanoDots may be outside the specified +/-5.5% accuracy range for radiation measurement. Landauer's investigation identified a potential non-conformance in the Optical Stimulated Luminescence (OSL) material used by the device to measure radiation exposure.
Users should contact Landauer to verify whether their dosimeters are affected and to obtain corrected or replacement units.
The recalled product
- Product
- Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SET; radiation monitoring dosimeter used with the microSTAR readers
- Manufacturer
- Landauer
- Hazard
- measurement-inaccuracy
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI/DI 0860003399903
- all batch numbers
Distribution
Distribution scope not specified by the agency.
Related recalls
Same category
- SevereMedtronic DLP Retrograde Cannula recalled for sterile barrier breach risk
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- SevereMedtronic DLP Retrograde Cannula Recalls for Sterile Barrier Breach Risk
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27