Radiation Dosimeter May Provide Inaccurate Dose Measurements
Landauer is recalling certain nanoDot D2DNS radiation dosimeters because they may measure radiation exposure outside the specified accuracy range. Approximately 5,958 units worldwide are affected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This Class II medical device may provide inaccurate radiation measurements. Although the potential non-conformance is not confirmed, the risk of measurement inaccuracy in a dosimeter creates a risk-of-harm scenario for workers relying on accurate exposure monitoring.
Plain-English summary
Landauer is recalling Calibrate (80 kVp) nanoDot D2DNS radiation monitoring dosimeters (Model Numbers 04217-000, 04217-KIT, and 04217-SET) used with microSTAR readers. Approximately 5,958 units have been distributed worldwide. The devices are designed to measure occupational radiation exposure.
Some nanoDots may be outside the specified +/-5.5% accuracy range for radiation measurement. Landauer's investigation identified a potential non-conformance in the Optical Stimulated Luminescence (OSL) material used by the device to measure radiation exposure.
Users should contact Landauer to verify whether their dosimeters are affected and to obtain corrected or replacement units.
The recalled product
- Product
- Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SET; radiation monitoring dosimeter used with the microSTAR readers
- Manufacturer
- Landauer
- Hazard
- Affected units
- 5,958
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI/DI 0860003399903
- all batch numbers
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 38 recalled by Landauer under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 38 products in this recall →Related recalls
Same manufacturer · Landauer
See all →- ModerateLandauer microStar Reader dosimetry system discontinued for medical applications
FDA (Devices) · 2024-03-27
- HighLandauer microSTARii reader recalled for measurement inaccuracy
FDA (Devices) · 2024-01-03
- HighLandauer nanoDot D2DXS Dosimeter Recall for Measurement Accuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot A Dosimeters Recalled for Potential Measurement Inaccuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot Radiation Dosimeters Recalled Due to Potential Accuracy Issues
FDA (Devices) · 2023-09-20
Same hazard · measurement inaccuracy
See all →- HighCDI OneView System BPM Probe temperature sensor measurement inaccuracy
FDA (Devices) · 2025-07-02
- ModeratePhilips Tempus Pro Patient Monitor Software Update for Operational Issues
FDA (Devices) · 2025-05-14
- SevereVentilator Model LD2110X23B Recalled for Potential Sensor Measurement Failure
FDA (Devices) · 2024-11-20
- SevereAeris EVO Ventilator May Develop Flow Sensor Fouling from Aerosols
FDA (Devices) · 2024-11-20
- SevereTrilogy EV300 ventilator: potential flow sensor malfunction from aerosol accumulation
FDA (Devices) · 2024-11-20
Similar recalls
- HighLandauer microSTARii reader recalled for measurement inaccuracy
FDA (Devices) · 2024-01-03
- ModerateLandauer microStar Reader dosimetry system discontinued for medical applications
FDA (Devices) · 2024-03-27
- HighPTW UNIDOS webline Dosimeters Recalled for Potential Radiation Over-Exposure
FDA (Devices) · 2015-09-02
- HighMobile X-Ray System Recalled for Software Calibration Error in Radiation Dose Meter
FDA (Devices) · 2022-10-05
- SevereRayStation 8B Software Recall for Dose Calculation Errors
FDA (Devices) · 2020-06-17
- HighDeWALT and Stanley Laser Distance Measurers Recalled for Excessive Radiation
FDA (Devices) · 2020-01-08