Landauer nanoDot D2DNS Radiation Dosimeters Recalled for Accuracy Non-Conformance
Landauer recalled 135 nanoDot D2DNS radiation dosimeters worldwide due to potential accuracy deviations outside the specified ±5.5% range in optical stimulated luminescence material.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II medical device recall involving potential measurement inaccuracy in radiation dosimeters. No injuries or illnesses are reported in the source documentation. The hazard is theoretical—a potential non-conformance in materials that may affect readings—placing it at High level as a risk-of-harm product without confirmed injury.
Plain-English summary
Landauer is recalling 135 nanoDot D2DNS radiation dosimeters (Model 03034-OTO), radiation monitoring devices distributed worldwide. These dosimeters measure occupational radiation exposure using optical stimulated luminescence (OSL) technology.
Investigation identified a potential non-conformance in the OSL material used in these dosimeters. This non-conformance may cause some units to read outside the specified accuracy range of ±5.5%. As a result, affected dosimeters may not accurately measure radiation exposure.
The recall affects occupational radiation workers and facilities worldwide that use these dosimeters for radiation monitoring.
The recalled product
- Product
- nanoDot D2DNS, for custom calibrate, OTO, Model Numbers 03034-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Manufacturer
- Landauer
- Hazard
- Affected units
- 135
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI/DI 0860003399903
- all batch numbers
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 38 recalled by Landauer under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 38 products in this recall →Related recalls
Same manufacturer · Landauer
See all →- ModerateLandauer microStar Reader dosimetry system discontinued for medical applications
FDA (Devices) · 2024-03-27
- HighLandauer microSTARii reader recalled for measurement inaccuracy
FDA (Devices) · 2024-01-03
- HighLandauer nanoDot D2DXS Dosimeter Recall for Measurement Accuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot A Dosimeters Recalled for Potential Measurement Inaccuracy
FDA (Devices) · 2023-09-20
- HighLandauer nanoDot Radiation Dosimeters Recalled Due to Potential Accuracy Issues
FDA (Devices) · 2023-09-20
Same hazard · measurement inaccuracy
See all →- HighCDI OneView System BPM Probe temperature sensor measurement inaccuracy
FDA (Devices) · 2025-07-02
- ModeratePhilips Tempus Pro Patient Monitor Software Update for Operational Issues
FDA (Devices) · 2025-05-14
- SevereVentilator Model LD2110X23B Recalled for Potential Sensor Measurement Failure
FDA (Devices) · 2024-11-20
- SevereAeris EVO Ventilator May Develop Flow Sensor Fouling from Aerosols
FDA (Devices) · 2024-11-20
- SevereTrilogy EV300 ventilator: potential flow sensor malfunction from aerosol accumulation
FDA (Devices) · 2024-11-20
Similar recalls
- HighLandauer microSTARii reader recalled for measurement inaccuracy
FDA (Devices) · 2024-01-03
- HighPTW UNIDOS webline Dosimeters Recalled for Potential Radiation Over-Exposure
FDA (Devices) · 2015-09-02
- HighAngiodynamics Recalls OARtrac PLUS Clinical Detector Units Due to Accuracy Issues
FDA (Devices) · 2019-12-25
- ModerateLandauer microStar Reader dosimetry system discontinued for medical applications
FDA (Devices) · 2024-03-27
- HighDeWALT and Stanley Laser Distance Measurers Recalled for Excessive Radiation
FDA (Devices) · 2020-01-08
- SevereRayStation 8B Software Recall for Dose Calculation Errors
FDA (Devices) · 2020-06-17