Tandem t:slim G4 Insulin Pump Recalled for Inaccurate Battery Readings
Tandem Diabetes Care is recalling 55 t:slim G4 Insulin Pumps because the fuel gauge may provide inaccurate battery readings, causing the pump to stop delivering insulin.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device that may stop delivering insulin, which is a risk-of-harm product. The source text does not report any specific injuries or illnesses, placing it in the 'High' category per the rubric.
Plain-English summary
Tandem Diabetes Care Inc is recalling 55 units of the t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM. The affected units are identified by specific serial numbers and are equipped with Software version 4.3.4.3 and Firmware version 004722.
The recall is due to a defect in the fuel gauge, the component that reads and reports battery parameters to the pump. This component may provide inaccurate readings, which can present to the user in two ways: by triggering a Malfunction 4 Alarm, or by triggering a succession of notifications (Alert 2, Alert 3, and Alarm 12) prior to the pump shutting off. In both scenarios, the alarms notify the user that the pump has stopped delivering insulin.
The affected products were distributed to consignees in the following U.S. states: IN, AZ, UT, TX, NY, NJ, MD, MN, IL, CO, AL, WI, CA, PA, OH, OR, MS, FL, NM, ID, NV, KY, and VA. Users of these specific serial numbers should contact Tandem Diabetes Care for instructions on how to proceed.
The recalled product
- Product
- t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
- Manufacturer
- Tandem Diabetes Care Inc
- Category
- Medical Device — Insulin Pump
- Affected units
- 55
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial Numbers: 483731
- 483930
- 483992
- 484013
- 492055
- 492368
- 501157
- 501177
- 501178
- 501187
- 501191
- 501200
- 501202
- 501205
- 501207
- 501268
- 501285
- 501364
- 501368
- 501370
Distribution
Related recalls
Same manufacturer · Tandem Diabetes Care Inc
See all →- HighTandem t:slim iOS App Recall Due to CGM Pairing Failure
FDA (Devices) · 2026-10-07
- HighTandem insulin pump Czech guide inverted a key connection instruction
FDA (Devices) · 2026-04-01
- CriticalInsulin pumps recalled over false vibration motor failure stopping delivery
FDA (Devices) · 2025-11-12
- HighInsulin pumps recalled over software defect causing unexpected correction bolus
FDA (Devices) · 2025-10-29
- HighInsulin pump app recalled over defect in right-to-left language settings
FDA (Devices) · 2025-10-01
Similar recalls
- HighInsulin pumps recalled for software malfunctions affecting blood glucose control
FDA (Devices) · 2022-10-05
- HighInsulin pumps recalled over software defect causing unexpected correction bolus
FDA (Devices) · 2025-10-29
- HighTandem t:slim X2 Insulin Pump Battery Display May Cause Insulin Delivery Failure
FDA (Devices) · 2022-06-22
- Severet:slim X2 Insulin Pump Software Defect Affecting Glucose Delivery
FDA (Devices) · 2025-04-23
- Hight:slim X2 Insulin Pumps Recalled for Software Issues Affecting Blood Sugar Management
FDA (Devices) · 2022-10-05
- CriticalInsulin pumps recalled over false vibration motor failure stopping delivery
FDA (Devices) · 2025-11-12