The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2457-2015·Announced 2015-09-02

RayStation 4.7.1 Radiation Therapy Software Recalled for Algorithm Error

RaySearch Laboratories is recalling RayStation 4.7.1 software due to an algorithm error that reverses contraction distances in radiation therapy planning.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a high-risk medical device software defect that could lead to incorrect treatment planning, but the source text does not report any specific injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

RaySearch Laboratories AB is recalling 126 programs of RayStation 4.7, version 4.7.1, a Radiation Therapy Treatment Planning System. The recall involves software build numbers 4.7.1.10 used with Structure Definition or Automatic Breast Planning features.

The recall is due to an issue with the algorithm for ROI (Region of Interest) contraction. When non-uniform contraction distances are used, the tool incorrectly interprets the input distances. Specifically, the right distance is interpreted as left, inferior as superior, and posterior as anterior. This error occurs because the six input distances are pairwise reversed.

The affected software was distributed in Colorado, Illinois, Michigan, North Carolina, Ohio, Tennessee, and Washington. Users of this specific software version should contact the manufacturer for further instructions on correcting the planning parameters.

The recalled product

Product
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.
Affected units
126

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • software build numbers are 4.7.1.10

Distribution

Distributed in 7 states:

Similar recalls