Arrow AC3 Optimus Intra-Aortic Balloon Pump Recalled for Vibration Failure
Arrow International is recalling AC3 Optimus Intra-Aortic Balloon Pumps because a component may fail due to vibration, potentially causing the device to stop working.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of…"
Plain-English summary
Arrow International Inc is recalling 2,123 units of the Arrow AC3 Optimus Intra-Aortic Balloon Pump (Product Code IAP-0700). The recall covers specific serial number ranges distributed worldwide.
The recall is due to a potential issue with a component within the device that is susceptible to vibration failure. This failure can result in fretting, charring, and discoloration of motor connector wires. These issues may cause the pump to display "System Error 3" and "High Baseline" alarms on the screen, leading to an abrupt cessation of function or an inability to start the IABP.
Healthcare providers and facilities should check their inventory for the affected serial numbers. The source text does not specify specific consumer actions beyond the recall notice, but the device should be evaluated for the described component issues.
The recalled product
- Product
- Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA, Product Code IAP-0700
- Manufacturer
- Arrow International Inc
- Category
- Medical Device — Cardiovascular
- Affected units
- 2,123
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 10 recalled by Arrow International Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 10 products in this recall →Related recalls
Same manufacturer · Arrow International Inc
See all →- HighArrow Arterial Kits Recalled Due to Open Packaging Seals
FDA (Devices) · 2024-10-02
- SevereIntra-Aortic Balloon Catheter Kit Recalled for Inflation and Structural Defects
FDA (Devices) · 2024-06-12
- SevereArrow UltraFlex Intra-Aortic Balloon Catheter Kit Recalled for Device Malfunction
FDA (Devices) · 2024-06-12
- SevereArrow Ultra 8 Intra-Aortic Balloon Catheter Kit Recalled for Inflation Defect
FDA (Devices) · 2024-06-12
- SevereARROW QuickFlash Radial Artery Catheters Recalled for Defective Component
FDA (Devices) · 2024-03-27
Same hazard · mechanical failure
See all →- SevereCanon Medical Alphenix INFX-8000C/B5 Recalled for Tabletop Shift Risk
FDA (Devices) · 2026-10-14
- ModerateLeMaitre TufTex Over-the-Wire Embolectomy Catheter Recalled for Balloon Failure
FDA (Devices) · 2026-10-14
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
Similar recalls
- SevereArrow AC3 Optimus Intra-Aortic Balloon Pump Battery Failure Recall
FDA (Devices) · 2022-12-21
- SevereArrow AC3 Optimus cardiac pump recalled due to battery issues
FDA (Devices) · 2022-12-21
- CriticalArrow AC3 Optimus IABP recalled for shortened battery run-times
FDA (Devices) · 2022-12-21
- SevereArrow AC3 Optimus Intra-Aortic Balloon Pumps recalled for short battery runtime
FDA (Devices) · 2022-12-21
- SevereArrow AC3 Optimus IABP Devices Recalled for Short Battery Run-Time Issue
FDA (Devices) · 2022-12-21
- SevereArrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Battery Run-Time Issue
FDA (Devices) · 2022-12-21