FDA (Devices) recall action 85684

Arrow International Inc recalled 10 products on 2020-07-08

The FDA (Devices) publishes a recall like this one as 10 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 10 pages.

Products
10
Announced
2020-07-08
Critical
0
Units
21,230

Why these products were recalled

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–10 of 10

  • SevereFDA (Devices)··2020-07-08

    Arrow AC3 Optimus Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International Inc is recalling AC3 Optimus Intra-Aortic Balloon Pumps because a component may fail due to vibration, potentially causing the device to stop working or fail to start.

    Product
    Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/AJLA, Product Code IAP-0701
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International is recalling 2,123 AutoCAT2 Intra-Aortic Balloon Pumps because a component may fail, causing the device to stop working or fail to start.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International Inc. is recalling 2,123 AutoCAT2 Intra-Aortic Balloon Pumps worldwide due to a component issue that may cause the device to stop working.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE DUTCH, Product Code IAP-0500NL
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International Inc. is recalling 2,123 AutoCAT2 Intra-Aortic Balloon Pumps because a component may fail due to vibration, potentially causing the device to stop working or fail to start.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE, Product Code IAP-0500
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Component Failure

    Arrow International Inc is recalling 2,123 AutoCAT2 Intra-Aortic Balloon Pumps worldwide due to a component issue that may cause the device to stop working.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International is recalling AutoCAT2 Intra-Aortic Balloon Pumps because a component may fail due to vibration, potentially causing the device to stop working.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE ITALIAN, Product Code IAP-0500I
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AC3 Optimus Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International is recalling AC3 Optimus Intra-Aortic Balloon Pumps because a component may fail due to vibration, potentially causing the device to stop working.

    Product
    Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA, Product Code IAP-0700
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International Inc is recalling 2,123 AutoCAT2 Intra-Aortic Balloon Pumps worldwide due to a component vibration issue that may cause the device to stop working.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Vibration Failure

    Arrow International is recalling 2,123 AutoCAT2 Intra-Aortic Balloon Pumps because a component may fail, causing the device to stop working or fail to start.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 SPANISH, Product Code IAP-0400E
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2020-07-08

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Component Vibration Failure

    Arrow International Inc. is recalling 2,123 AutoCAT2 Intra-Aortic Balloon Pumps because a component vibration issue may cause the device to stop working or fail to start.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE SPANISH, Product Code IAP-0500E
    Category
    Distribution
    0 states