Infusomat SPACE IV Pump Set Recalled for Backcheck Valve Malfunction
B. Braun is recalling Infusomat SPACE IV pump sets due to a backcheck valve malfunction that could cause medication to flow backward into the primary IV container and prevent proper priming of the line.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device with potential for serious patient harm (adverse drug reactions, loss of critical medication) due to backcheck valve malfunction. No illnesses or injuries reported in the source material, placing this within the risk-of-harm category without confirmed reported harm.
Plain-English summary
B. Braun Medical, Inc. is recalling the Infusomat SPACE IV Pump Set with 2 Caresite, Model 490103. The recall includes 643,608 units with UDI-DI 04046964302107, affecting all lots distributed after August 1, 2023. Distribution was worldwide, including throughout the US and Canada.
The backcheck valve in these IV sets can malfunction, resulting in medication flowing backward from secondary (piggyback) IV containers into the primary IV container. The malfunction may also prevent the IV line from being primed correctly.
These failures could cause adverse drug reactions and loss of medication. Patients receiving IV medication through affected units may be at risk.
The recalled product
- Product
- Infusomat SPACE PUMP IV SET W/2 CARESITE, CKVLV -Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 490103
- Manufacturer
- B. Braun Medical, Inc.
- Affected units
- 643,608
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI-DI 04046964302107
- All Lots Distributed After 01AUG2023
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 29 recalled by B. Braun Medical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 29 products in this recall →Related recalls
Same manufacturer · B. Braun Medical, Inc.
See all →- ModerateB Braun SAFSITE Injection Site Valve Recall for Leakage Risk
FDA (Devices) · 2026-10-28
- CriticalB. Braun Recalls Sodium Chloride Injection Due to Particulate Matter
FDA (Drugs) · 2026-09-16
- LowB Braun Recalls Discofix Stopcock Sets Due to Incorrect Expiration Date
FDA (Devices) · 2026-09-09
- SevereB. Braun Recalls Lactated Ringer's Injection Due to Particulate Matter
FDA (Drugs) · 2026-09-02
- ModerateB. Braun kits contain Mastisol adhesive with potentially cracking tubing
FDA (Devices) · 2026-06-10
Same hazard · valve malfunction
See all →- ModerateNitrous Oxide Cartridges Recalled Due to Premature Valve Opening
FDA (Devices) · 2025-02-12
- HighInfusion Pump Set Recalled for Potential Backcheck Valve Malfunction and Medication Backflow
FDA (Devices) · 2024-07-24
- HighIV Infusion Pump Sets Recalled for Backcheck Valve Malfunction
FDA (Devices) · 2024-07-24
- HighInfusomat SPACE Pump IV Sets Recalled for Backcheck Valve Malfunction
FDA (Devices) · 2024-07-24
- HighInfusomat SPACE IV infusion sets recalled for backcheck valve malfunction
FDA (Devices) · 2024-07-24
Similar recalls
- HighIV set 490067 recalled over medication backflow between IV containers
FDA (Devices) · 2025-12-03
- HighB. Braun Recalls Infusomat Space Pump IV Sets Due to Missing Filter
FDA (Devices) · 2018-07-04
- HighIV set 490542 recalled over backflow between primary and secondary containers
FDA (Devices) · 2025-12-03
- HighIV set 490612 recalled over medication backflow between IV containers
FDA (Devices) · 2025-12-03
- HighIV set 490408 recalled over backflow between primary and secondary containers
FDA (Devices) · 2025-12-03
- HighIV set 490385 recalled over medication backflow between IV containers
FDA (Devices) · 2025-12-03