The Recall Desk

FDA (Devices) recall action 94823

B. Braun Medical, Inc. recalled 29 products on 2024-07-24

The FDA (Devices) publishes a recall like this one as 29 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 29 pages.

Products
29
Announced
2024-07-24
Critical
0
Units
17,981,522

Why these products were recalled

Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 29

  • HighFDA (Devices)··2024-07-24

    Infusomat IV Pump Set Recalled for Backcheck Valve Malfunction

    The Infusomat 60DROP METRISET PUMP SET may experience backcheck valve malfunction, causing medication to flow backward between IV containers and loss of medication. This could result in adverse drug reactions or inability to administer medication properly.

    Product
    Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362034
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Infusion Pump Sets Recalled for Backcheck Valve Malfunction

    B. Braun is recalling Infusomat IV infusion pump sets due to potential backcheck valve malfunction that can cause medication to flow backward between IV containers and prevent proper medication delivery.

    Product
    Infusomat UNIV. 15 DROP PUMP SET W/2 CARESITE CV- Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number 490102
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusion Pump Set Recalled for Potential Backcheck Valve Malfunction and Medication Backflow

    B. Braun's Infusomat UNIV. 15 IV infusion pump set is being recalled due to a potential backcheck valve malfunction that could cause medication to backflow from secondary into primary IV containers, potentially resulting in patient injury.

    Product
    Infusomat UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363030
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusomat UNIV 60 IV Infusion Pump Set Recalled for Valve Malfunction Risk

    B. Braun is recalling Infusomat UNIV 60 IV infusion pump sets due to a backcheck valve defect that could cause medication backflow between IV containers and prevent priming, potentially resulting in adverse drug reactions or medication loss.

    Product
    Infusomat UNIV 60 DROP PUMP SET, W/2 CARESITE INJ. -Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number : 490104
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Infusion Pump Set Valve May Cause Dangerous Medication Backflow

    B. Braun's Infusomat pump sets may have faulty backcheck valves causing medication to backflow between IV bags, risking reduced medication delivery and adverse reactions. All units distributed worldwide after August 2023 are affected.

    Product
    Infusomat UNIV. 15 DROP PUMP SET W 3 CARESITE L.L. Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 490100
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Infusion Pump Set Backcheck Valve Malfunction Recall

    B. Braun Medical recalls Infusomat UNIV. 15 DROP PUMP SET due to backcheck valve malfunction that could cause medication backflow and loss of medication, resulting in adverse drug reactions.

    Product
    Infusomat UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363433
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusomat UNIV. 15 IV Pump Set Recalled for Backflow Hazard

    B. Braun Medical is recalling 32,752 units of the Infusomat UNIV. 15 DR pump tubing set due to potential backcheck valve malfunction that could cause medication backflow between IV containers and medication loss.

    Product
    Infusomat UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363424
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Outlook IV Infusion Pump Set Recalled Due to Backcheck Valve Malfunction Risk

    B. Braun is recalling Outlook IV pump sets due to potential backcheck valve malfunction that could allow medication backflow and prevent proper priming, posing a risk of adverse drug reactions and medication loss.

    Product
    Outlook OUTLOOK PUMP SET,3 CARESITE LADS,115 IN. Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 490361
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    B. Braun Infusomat 60 IV Pump Sets Recalled for Backcheck Valve Malfunction

    B. Braun is recalling Infusomat 60 infusion pump sets (Model 363010) due to potential backcheck valve malfunction. The defect could cause medication to flow backward into primary IV containers and prevent proper medication delivery.

    Product
    Infusomat 60 DROP METRISET PUMP SET,3 SFLINE ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363010
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusion Pump Sets Recalled for Potential Backflow of Medication

    B. Braun is recalling Infusomat UNIV. 15 IV pump sets due to a backcheck valve malfunction that could divert medication between containers and prevent proper priming.

    Product
    Infusomat UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication. Model/Catalog Number: 362033
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Pump Sets Recalled for Potential Backcheck Valve Malfunction

    B. Braun Infusomat SPACE IV pump sets may experience backcheck valve malfunction, causing medication backflow between containers and preventing proper IV line priming. This could result in adverse drug reactions, incorrect medication doses, or medication loss.

    Product
    Infusomat SPACE PUMP IV SET , 120 IN.- Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362432
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    B. Braun Infusomat SPACE IV Set Backcheck Valve Malfunction Recall

    B. Braun Medical is recalling 68,205 Infusomat SPACE IV Sets due to a potential backcheck valve malfunction that could allow medication backflow between IV containers.

    Product
    Infusomat SPACE PUMP IV SET, 2 CARESITES, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363420
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Infusion Set Backcheck Valve Malfunction May Cause Medication Loss

    B. Braun is recalling OUTLOOK IV Sets due to potential backcheck valve malfunction that could allow medication backflow and loss. The defect may result in adverse drug reactions or medication loss.

    Product
    OUTLOOK IV SET 15DROP W/3 CARESITE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication. Model/Catalog Number: 354213
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusomat UNIV. 15 IV Pump Set Recalls Due to Backcheck Valve Malfunction

    B. Braun is recalling Infusomat UNIV. 15 IV pump sets because the backcheck valve may malfunction, potentially causing medication backflow or loss, which could lead to adverse drug reactions.

    Product
    Infusomat UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363430
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusomat SPACE IV Pump Set Recalled for Backcheck Valve Malfunction

    B. Braun is recalling Infusomat SPACE IV pump sets due to a backcheck valve malfunction that could cause medication to flow backward into the primary IV container and prevent proper priming of the line.

    Product
    Infusomat SPACE PUMP IV SET W/2 CARESITE, CKVLV -Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 490103
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusion pump set backcheck valve malfunction may prevent medication delivery

    B. Braun Medical is recalling an infusion pump set due to potential backcheck valve malfunction that could cause medication to flow backward into IV containers and prevent proper medication administration. The recall affects 3,624 units distributed worldwide.

    Product
    Infusomat UNIV. 15 DROP PUMP SET W/ 0.2 FILTER- Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362050
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Pump Infusion Set Backcheck Valve May Malfunction and Cause Medication Backflow

    B. Braun is recalling the Infusomat UNIV.15DROP PUMP SET (Model 362032) because the backcheck valve may malfunction, allowing medication to flow backward from secondary to primary IV containers, risking medication loss and adverse drug reactions.

    Product
    Infusomat UNIV.15DROP PUMP SET W/3 SAFELINE LL-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362032
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV infusion set recalled due to backcheck valve malfunction

    B. Braun is recalling OUTLOOK IV sets due to potential backcheck valve malfunction that could allow medication to flow backward between IV containers and prevent proper priming, potentially resulting in patient injury.

    Product
    OUTLOOK IV SET 15DROP W/2 CARESITE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication. Model/Catalog Number: 354212
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Infusion Pump Set Recalled for Backcheck Valve Malfunction Risk

    B. Braun is recalling the Infusomat UNIV. 15 DROP PUMP SET due to potential backcheck valve malfunction that could cause medication backflow and loss. The defect may result in adverse drug reactions or improper medication delivery.

    Product
    Infusomat UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 363230
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusion pump IV administration set recalled for backcheck valve malfunction risk

    B. Braun is recalling Infusomat UNIV. 15 DROP PUMP SET infusion sets due to potential backcheck valve malfunction that may cause medication to flow backward into primary IV containers and prevent proper priming, risking adverse drug reactions.

    Product
    Infusomat UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363421
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Infusion Pump Set Recalled for Medication Backflow Risk

    B. Braun's Infusomat 60 IV infusion pump set is being recalled due to a backcheck valve that may malfunction, allowing medication to flow backward between IV containers and preventing proper priming of the line.

    Product
    Infusomat 60 DROP METRISET PUMP SET, 3 CARESITES Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 490101
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusion Pump Component Valve Malfunction May Cause Medication Backflow

    B. Braun's Infusomat infusion pump component can malfunction, causing medication to flow backward from secondary IV lines into primary lines, potentially resulting in improper medication delivery and patient harm.

    Product
    Infusomat 60DROP METRISET PUMP, 3 CARESITES, ASV-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 363422
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    Infusomat SPACE Pump IV Sets Recalled for Backcheck Valve Malfunction

    B. Braun's Infusomat SPACE Pump IV sets are recalled due to a potential backcheck valve malfunction that could cause medication backflow and inability to prime. The defect affects approximately 109,464 units distributed in the US and Canada.

    Product
    Infusomat SPACE PUMP IV SET 15D, 110 IN. -Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 480254
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    B. Braun Infusomat IV Pump Set Recalled Due to Backflow Risk

    B. Braun Medical recalls Infusomat UNIV. 15 DROP PUMP SET units due to a potential backcheck valve malfunction that could cause medication backflow and adverse drug reactions.

    Product
    Infusomat UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362431
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-07-24

    IV Infusion Pump Set Recalled for Backcheck Valve Malfunction Risk

    B. Braun is recalling Infusomat SPACE PUMP SET 15 DROP W/2 SAFEDAY (Model 362420) IV infusion sets due to backcheck valve malfunction that can cause medication to flow backward from secondary to primary IV containers.

    Product
    Infusomat SPACE PUMP SET 15 DROP W/2 SAFEDAY-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362420
    Category
    Distribution
    0 states