The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2371-2018·Announced 2018-07-11

Zimmer Biomet Recalls M/L Taper Hip Prostheses Due to Black Residue

Zimmer Biomet is recalling specific M/L Taper Hip Prostheses manufactured before October 2015 because black residue may be present in the packaging.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a high-risk medical device where the hazard (debris causing wear) presents a risk of harm, but the source text does not report any specific illnesses or injuries, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

Zimmer Biomet, Inc. is recalling 182 units of the Zimmer M/L Taper Hip Prosthesis With Kinective Technology Cementless, Size 17.5 (catalog number 00-7713-017-00). The recall covers devices manufactured prior to October 2015. According to the source text, these products were distributed solely to Japan.

The recall is due to the potential for black residue to be observed within the polyethylene bag containing the implant. The firm states that this residue is a combination of the titanium plasma spray used to create the porous surface of the implant and the polyethylene bag. The recalling firm asserts that the residue is biocompatible.

The identified risks include the potential delay of surgery while locating a replacement device and the possibility of surgical intervention due to wear caused by debris. No specific illnesses or injuries are reported in the source text.

The recalled product

Product
Zimmer¿ M/L Taper Hip Prosthesis With Kinective¿ Technology Cementless, Size 17.5; 00-7713-017-00
Affected units
182

Distribution

Distribution scope not specified by the agency.

Part of a larger recall action

This product is one of 15 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 15 products in this recall →