Stryker Restoration Anatomic Shell Recalled for Packaging Defect
Howmedica Osteonics is recalling Stryker Restoration Anatomic Hemispherical Shells because the device may fall off its post within the inner packaging, resulting in loose coating debris.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, which typically corresponds to a Moderate severity score when no specific injuries or hospitalizations are reported in the source text. The hazard is a packaging defect leading to potential debris, which is a lower risk than structural failure or contamination.
Plain-English summary
Howmedica Osteonics Corp. is recalling 4,895 units of the Stryker Restoration Anatomic Hemispherical Shell, Size 56 Right, Sterile (Catalog # 504-02-56D-R). The recall affects specific lot numbers, including EE5WWV, DY6V66, and 4T75EE, among others. The devices were distributed worldwide, including in numerous U.S. states such as New York, Texas, and California.
The recall was initiated because the Stryker Restoration Anatomic Shell may fall off its post within its inner packaging. This defect can result in the device being loose in the package with coating debris. The FDA has classified this recall as Class II, indicating that use of, or exposure to, the violative product may cause temporary or medically reversible adverse health consequences.
Healthcare providers and facilities that have received these devices should stop using them and contact Howmedica Osteonics Corp. for instructions on return or replacement. Patients who have had these devices implanted should contact their healthcare provider to discuss the recall and any necessary follow-up care.
The recalled product
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 56 Right, Sterile, Catalog # 504-02-56D-R, UDI (01)07613327012347(17)
- Manufacturer
- Howmedica Osteonics Corp.
- Hazard
- Affected units
- 4,895
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Lot #s EE5WWV
- DY6V66
- 4T75EE
- HT3X47
- YX8KLL
- 99554T
- T58TR0
- A46AJA
- P23W2H
- 9X0WHV
- 681E6V
- XP0AKD
- RD53PW
- R67PR4
- N11PLW
- WW20KE
- KV6P9X
- 6X4R9T
- 2X4KLD
- M08065
Distribution
Part of a larger recall action
This product is one of 16 recalled by Howmedica Osteonics Corp. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 16 products in this recall →Related recalls
Same manufacturer · Howmedica Osteonics Corp.
See all →- SevereHowmedica Recalls Scorpio Patella Implants Due to Polyethylene Wear Risk
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Orthopaedic Implants Due to Potential UHMWPE Oxidation
FDA (Devices) · 2026-09-23
- SevereHowmedica Recalls Orthopaedic Implants Due to Potential UHMWPE Oxidation
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Orthopaedic Implant Components Due to Material Aging
FDA (Devices) · 2026-09-23
- HighHowmedica Recalls Scorpio Receded Metal Backed Patella Implants
FDA (Devices) · 2026-09-23
Same hazard · packaging defect
See all →- HighEmbosphere Microspheres Recalled Due to Packaging Sterility Concerns
FDA (Devices) · 2026-10-07
- ModerateAmneal Recalls Estradiol Gel Packets Due to Empty or Partially Full Containers
FDA (Drugs) · 2026-10-07
- HighCardinal Health Recalls Presource Convenience Kits Due to Seal Defects
FDA (Devices) · 2026-09-30
- ModerateAVID Medical Recalls Halyard IV Start Kits Due to Packaging Seal Issues
FDA (Devices) · 2026-09-30
- HighCardinal Health Recalls Presource Convenience Kits Due to Potential Seal Defects
FDA (Devices) · 2026-09-30
Similar recalls
- ModerateStryker Howmedica Hip Prosthesis Components Recalled for Seal Defects
FDA (Devices) · 2015-05-27
- ModerateStryker Triathlon Femoral Distal Augment Recalled for Packaging Label Errors
FDA (Devices) · 2013-05-29
- ModerateStryker Recalls rHead and uHead Implants Due to Packaging Integrity Concerns
FDA (Devices) · 2015-08-26
- HighHowmedica Recalls Stryker Radial Head Implants Due to Revision Rate
FDA (Devices) · 2018-08-15
- ModerateStryker Triathlon Femoral Component Recalled for Sterile Barrier Seal Issues
FDA (Devices) · 2015-05-27
- SevereHowmedica Recalls rHead Recon Stem Implant Due to Higher Revision Rates
FDA (Devices) · 2018-08-15