St Jude Medical Recalls Quadra Assura MP Implantable Cardiac Devices for Security Update

St Jude Medical is recalling 68,117 Quadra Assura MP implantable cardiac devices to release new firmware that strengthens security against unauthorized access.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (implantable cardiac device) with a security vulnerability that could allow unauthorized access, but no injuries or illnesses have been reported.

Plain-English summary

St Jude Medical Inc. is recalling 68,117 units of the Quadra Assura MP, Sterile EO, implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D). The recall is being conducted to release new firmware intended to strengthen security by providing an additional layer of protection against unauthorized device access.

The affected devices were distributed nationwide in the United States. For older generation devices that are not capable of accepting the new firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.

The recalled product

Product
Quadra Assura MP, Sterile EO, Model #/ Part #: CD3269-40/100043177; CD3269-40Q/100043178; CD3271-40/100043130; CD3271-40Q/100043191; CD3369-40C/100080339, 100080478; CD3369-40Q/100080439, 100080476, 100105528; CD3371-40/100078814, 100079232, 100079513…
Affected units
68,117

Distribution

Distributed nationwide across the United States.