GE Allia IGS 3 Pulse systems pose shock risk to servicers
Allia IGS 3 Pulse angiographic X-ray systems are being recalled because an electrical cable issue could expose service personnel to electrical shock while power is on. Nine units affected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall with a stated electrical shock risk to service personnel, a risk of harm with no injuries reported in the source.
Plain-English summary
The Allia IGS 3 Pulse angiographic X-ray system, UDI-DI 00195278719577, is being recalled. The action covers nine units, identified by the serial numbers in the recall notice including M3-24-053, M3-24-124 and M3-25-029.
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. According to the manufacturer, this issue could lead to a risk of electrical shock to service personnel performing installation, maintenance or troubleshooting activities inside the C-FRT cabinet while power is on. The FDA classified the action as Class II.
Distribution was worldwide, including nationwide in the United States. The systems are installed in healthcare facilities rather than sold to consumers.
Facilities should identify affected systems using the UDI-DI and serial numbers in the recall notice and follow the instructions issued by GE HealthCare.
The recalled product
- Product
- Allia IGS 3 Pulse angiographic X-ray system
- Manufacturer
- GE Medical Systems, LLC
- Affected units
- 9
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (11)
- UDI-DI: 00195278719577
- Serial Numbers: M3-24-053
- M3-24-124
- M3-25-029
- M3-25-051
- M3-25-053
- M3-25-091
- M3-25-098
- M3-25-124
- M3-26-010
- M3-26-030
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 5 recalled by GE Medical Systems, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · GE Medical Systems, LLC
See all →- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- HighGE Medical Optima XR200/220/240amx Mobile Radiographic Systems Recall
FDA (Devices) · 2026-07-22
- HighGiraffe OmniBed incubator canopy fasteners can fall into patient compartment
FDA (Devices) · 2026-06-24
- HighGiraffe OmniBed replaceable units share the canopy fastener disengagement defect
FDA (Devices) · 2026-06-24
Same hazard · electrical shock
See all →- High
- Severe
- HighFort Defiance Automated Field Steam Sterilizer Recalled for Electrical Grounding Defect
FDA (Devices) · 2026-08-26
- SevereJayco Recalls 2026 Seismic Fifth Wheel Toy Haulers Due to Fire Risk
NHTSA · 2026-07-02
- Severe
Similar recalls
- HighGE Healthcare X-ray Systems: Potential Adhesive Failure Poses Fall Hazard
FDA (Devices) · 2024-05-15
- HighGE Healthcare Allia IGS 7 X-ray Systems Adhesive Bolt Failure Recall
FDA (Devices) · 2024-05-15
- HighGE Healthcare Allia IGS Fluoroscopic Systems Recall: Missing Adhesive
FDA (Devices) · 2024-05-15
- HighGE Innova IGS Fluoroscopic X-Ray Systems Adhesive Defect Recall
FDA (Devices) · 2024-05-15
- HighGE Allia IGS Fluoroscopic X-Ray Systems Adhesive Bolt Defect
FDA (Devices) · 2024-05-15
- HighGE Healthcare X-Ray Systems Recalled for Loose Bolts on Gantry Assembly
FDA (Devices) · 2024-05-15