The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2310-2015·Announced 2015-08-05

Anspach Diamond Ball Dissection Tools Recalled for Inconsistent Shaft Exposure

The Anspach Effort, Inc. is recalling 70 Anspach 5MM Diamond Ball dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device used for cutting bone where the physical exposure of the tool may be greater than labeled, posing a risk of harm to the patient or user, although no specific injuries are reported in the source text.

Plain-English summary

The Anspach Effort, Inc. is recalling 70 units of the ANSPACH 5MM Diamond Ball, 2MM Shaft Exposure (REF S-5D-2-G1). These tools are used for cutting and shaping bone, including the spine and cranium. The recall covers all lot numbers of this specific reference number.

The recall was initiated because the G1 dissection tools are not compatible with all attachments listed on the labeling. The tools are labeled with a specified shaft exposure, but this exposure may not be consistent when used with each attachment listed on the product label. Specifically, the shaft exposure may be greater than the labeled amount.

The affected tools were distributed worldwide, including in the United States (AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX) and internationally to Canada and China. Users should stop using the affected tools and contact the recalling firm for instructions on return or replacement.

The recalled product

Product
ANSPACH - 5MM Diamond Ball, 2MM Shaft Exposure; REF S-5D-2-G1 Cutting shaping bone including spine and cranium.
Affected units
70

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • REF S-5D-2-G1
  • ALL lot numbers.

Distribution

Distributed in 17 states:

Part of a larger recall action

This product is one of 34 recalled by The Anspach Effort, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 34 products in this recall →

Similar recalls