The Recall Desk

FDA (Devices) recall action 71212

The Anspach Effort, Inc. recalled 34 products on 2015-08-05

The FDA (Devices) publishes a recall like this one as 34 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 34 pages.

Products
34
Announced
2015-08-05
Critical
0
Units
5,320

Why these products were recalled

G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–25 of 34

  • HighFDA (Devices)··2015-08-05

    Anspach Diamond Ball Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 72 Anspach Extra Coarse Diamond Ball tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure.

    Product
    ANSPACH - Extra Coarse Diamond Ball; 8MM REF L-8DX-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach Effort Recalls G1 Dissection Tools Due to Labeling Inconsistencies

    The Anspach Effort, Inc. is recalling 796 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure than specified.

    Product
    ANSPACH -4MM Carbide Fluted Ball, Extends 27MM; REF S-4BL-C-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 686 G1 dissection tools because the specified shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 3MM Fluted Matchstick, Extends 8.9MM, Less Aggressive; REF S-8NS-M-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach 7mm Diamond Ball Dissection Tools Recalled for Labeling Inconsistency

    The Anspach Effort, Inc. is recalling 75 Anspach 7mm Extra Coarse Diamond Ball dissection tools because the specified shaft exposure may be inconsistent with the labeled attachments.

    Product
    ANSPACH - 7MM Extra Coarse Diamond Ball; REF M-7DX-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach Fluted Drum Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 65 Fluted Drum dissection tools because the specified shaft exposure may be inconsistent with attachments, potentially resulting in greater exposure.

    Product
    ANSPACH - Fluted Drum, Side Cut only 10MM x 14MM , REF L-10DRC-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 88 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure than specified.

    Product
    ANSPACH - 4MM Fluted Ball, Extends 27MM; REF M-4B-27-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 80 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure than specified.

    Product
    ANSPACH - 5MM Coarse Diamond Ball, 22MM Shaft Exposure; REF S-5DC-22-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • HighFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 70 G1 dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 6MM Coarse Diamond Ball, 2 mm Shaft Exposure; REF M-6DC-2-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 95 G1 dissection tools because the specified shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 8MM x 12.5MM Fluted Barrel; REF M-8BA-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach Fluted Barrel Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 110 Anspach Fluted Barrel dissection tools because the specified shaft exposure may be inconsistent with attachments, potentially resulting in greater exposure.

    Product
    ANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • HighFDA (Devices)··2015-08-05

    Anspach Diamond Ball Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 70 Anspach 5MM Diamond Ball dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 5MM Diamond Ball, 2MM Shaft Exposure; REF S-5D-2-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach Fluted Drum Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 115 fluted drum dissection tools because the specified shaft exposure may be inconsistent with certain attachments, potentially resulting in greater exposure than labeled.

    Product
    ANSPACH - Fluted Drum, Side Cut Only; 7MM x 14MM REF L-7DRC-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach Fluted Acorn Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 227 Fluted Acorn dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - Fluted Acorn, 7MM REF L-7AC-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • HighFDA (Devices)··2015-08-05

    Anspach Fluted Drum Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 177 fluted drum dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - Fluted Drum, Side Cut only, 6MM x 14MM REF L-6DRC-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 92 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure.

    Product
    ANSPACH - 4MM x 11 MM Fluted Barrel, 14 Flutes; REF M-4BA-A-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 118 G1 dissection tools because the specified shaft exposure on the label may not be consistent with all listed attachments.

    Product
    ANSPACH - 5MM Fluted Ball, Extends 27MM; REF M-5B-27-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 91 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure than specified.

    Product
    ANSPACH - 8MM x 14MM Fluted Drum, Side Cut Only; REF L-8DRC-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • HighFDA (Devices)··2015-08-05

    Anspach Fluted Drum Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 138 fluted drum dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - Fluted Drum, Side Cut only 5MM x 14MM , REF L-5DRC-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • HighFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 99 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual usage, potentially resulting in greater exposure than specified.

    Product
    ANSPACH -4MM Coarse Diamond Ball, 22MM Shaft Exposure; REF S-4DC-22-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 69 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure.

    Product
    ANSPACH - Fluted Drum, Side Cut Only; 9MM x 14 MM REF L-9DRC-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach Extra Coarse Diamond Drum Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 88 Extra Coarse Diamond Drums because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • HighFDA (Devices)··2015-08-05

    Anspach 5MM Fluted Ball Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 79 Anspach 5MM Fluted Ball dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 5MM Fluted Ball, 2 mm Shaft Exposure; REF M-5B-2-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 78 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure.

    Product
    ANSPACH -1.5MM x 7.5MM Cylindrical Carbide Drum, Extends 25MM; REF S-15DRMC-L-G1. Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach Fluted Barrel Dissection Tools Recalled for Labeling Inconsistency

    The Anspach Effort, Inc. is recalling 110 fluted barrel dissection tools because the labeled shaft exposure may be inconsistent with actual usage.

    Product
    ANSPACH - Fluted Barrel; 8MM x 12.5MM REF L-8BA-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states
  • ModerateFDA (Devices)··2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 95 G1 dissection tools because the labeled shaft exposure may not be consistent with all listed attachments, potentially resulting in greater exposure than specified.

    Product
    ANSPACH - 6MM x 11.3MM Fluted Barrel; REF M-6BA-A-G1 Cutting shaping bone including spine and cranium.
    Category
    Distribution
    17 states