Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling
The Anspach Effort, Inc. is recalling 95 G1 dissection tools because the specified shaft exposure may be greater than labeled when used with certain attachments.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a labeling inconsistency regarding shaft exposure without reported injuries or illnesses, fitting the criteria for a moderate severity rating.
Plain-English summary
The Anspach Effort, Inc. is recalling 95 units of the ANSPACH 8MM x 12.5MM Fluted Barrel (REF M-8BA-G1) dissection tools. These tools are used for cutting and shaping bone, including the spine and cranium.
The recall is due to a labeling inconsistency where the G1 dissection tools are not compatible with all attachments listed on the labeling. Specifically, the tools are labeled with a Specified Shaft Exposure that may not be consistent when used with each attachment listed on the product label, potentially resulting in greater shaft exposure than indicated.
The recalled products were distributed worldwide, including in the United States (AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX) and internationally to Canada and China. Users should stop using the affected tools and contact the manufacturer for instructions.
The recalled product
- Product
- ANSPACH - 8MM x 12.5MM Fluted Barrel; REF M-8BA-G1 Cutting shaping bone including spine and cranium.
- Manufacturer
- The Anspach Effort, Inc.
- Affected units
- 95
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- REF M-8BA-G1
- ALL lot numbers
Distribution
Part of a larger recall action
This product is one of 34 recalled by The Anspach Effort, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 34 products in this recall →Related recalls
Same manufacturer · The Anspach Effort, Inc.
See all →- HighAdult Craniotome Bearings May Detach During Surgical Use
FDA (Devices) · 2023-01-18
- HighCraniotome Attachment Ball Bearings May Separate During Surgery
FDA (Devices) · 2023-01-18
- HighLarge Craniotome Attachment Ball Bearings Risk Loosening During Surgery
FDA (Devices) · 2023-01-18
- HighAnspach Adult Rotating Craniotome Attachment ball bearings may detach during surgery
FDA (Devices) · 2023-01-18
- HighCraniotome Attachment Ball Bearings May Detach During Surgery
FDA (Devices) · 2023-01-18
Same hazard · mis labeling
See all →- ModerateMedline Trach Tube Kit Recalled for Component Labeling Discrepancy
FDA (Devices) · 2026-09-02
- ModerateAVID Medical Recalls Halyard Total Joint Kits for Latex Labeling Deficiency
FDA (Devices) · 2026-09-02
- ModerateOlympus Recalls PowerSeal Sealer Due to Incorrect Outer Box Labeling
FDA (Devices) · 2026-09-02
- HighSura Tanmen Ramen Noodles Recalled for Undeclared Fish Allergens
FDA (Food) · 2026-09-02
- SevereWhite Vein Advanced Alkaloids Chewable Tablets Recall for Excess Active Ingredient
FDA (Food) · 2026-09-02