HLD Systems Washer Pasteurizer Recalled for Terminal Block Overheating
Cenorin, LLC is recalling 14 HLD Systems 600 Series Washer/Pasteurizers because a supplier quality issue with terminal blocks may cause overheating, charring, or burning.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (fire/burn) where injury has not yet been reported, which aligns with the High severity criteria.
Plain-English summary
Cenorin, LLC is recalling 14 units of the HLD Systems 600 Series Washer/Pasteurizer, specifically Model 610HT in 115V and 230V configurations. The recall is due to a supplier quality issue with the terminal blocks, which may cause the components to overheat, char, and/or burn.
The affected units were distributed worldwide, including nationwide in the United States in the states of California, Florida, Indiana, North Carolina, Nebraska, Utah, and West Virginia, as well as in Malaysia and Taiwan. The specific serial numbers for the recalled units are 611503, 611504, 611505, 611506, 611601, 611602, 611603, 611604, 611605, 611606, 611607, 611608, 611609, and 611610.
Users of these devices should contact Cenorin, LLC for further instructions on how to address the potential safety hazard.
The recalled product
- Product
- HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V Product Usage: Clean and thermally disinfect medical devices using full immersion pasteurization
- Manufacturer
- Cenorin, LLC
- Hazard
- Affected units
- 14
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (13)
- Serial Numbers: 611503
- 611504
- 611505
- 611506
- 611601
- 611602
- 611603
- 611604
- 611605
- 611606
- 611607
- 611608
- 611609 & 611610
Distribution
Related recalls
Same manufacturer · Cenorin, LLC
See all →- HighCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- HighCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- SevereCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- HighCenorin HLD System Model 610 Recalled for Temperature Sensor Defect
FDA (Devices) · 2014-07-16
- ModerateCenorin HLD Systems Model 610 Washer Recalled for Cleaning Fluid Defect
FDA (Devices) · 2014-06-18
Same hazard · fire
See all →- SevereVQQQ Gas Patio Heaters Recalled for Fire and Burn Hazards
CPSC · 2026-10-22
- HighAutocar Recalls Xpeditor Trucks for Fuel Hose Separation Risk
NHTSA · 2026-10-21
- SevereAudi Recalls 2024-2025 Q4 E-Tron Models Due to Battery Fire Risk
NHTSA · 2026-10-21
- HighForest River Recalls 2023-2026 Pause Travel Trailers for Wiring Defect
NHTSA · 2026-10-20
- SevereForest River Recalls 2026-2027 Cedar Creek Trailers for Fire Risk
NHTSA · 2026-10-16
Similar recalls
- HighCenorin HLD System Model 610 Recalled for Temperature Sensor Defect
FDA (Devices) · 2014-07-16
- SevereCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- ModerateCenorin HLD Systems Model 610 Washer Recalled for Cleaning Fluid Defect
FDA (Devices) · 2014-06-18
- HighReliance Synergy Washer/Disinfector fire hazard due to electrical contactor malfunction
FDA (Devices) · 2022-06-15
- HighWasher disinfectors recalled over overheating risk from incorrect pressure calibration
FDA (Devices) · 2025-10-08
- LowAdvanced Sterilization Products Recalls STERRAD 200 Sterilizers for Odor Emission
FDA (Devices) · 2013-10-30