Cenorin HLD System Model 610 Recalled for Temperature Sensor Defect
Cenorin is recalling 25 HLD System Model 610 units because a temperature sensor defect may prevent water from heating during pasteurization.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a failure to heat water for pasteurization, which is a risk-of-harm product defect where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Cenorin, LLC is recalling 25 units of the HLD System, Model 610, a medical device cleaning and high-level disinfection washer/pasteurizer. The recall was initiated because the temperature sensor/control system may provide a false temperature reading. This defect caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.
The affected units were distributed worldwide, including in the United States (California, Delaware, Florida, Massachusetts, Michigan, New York, Pennsylvania, Texas, Utah, Washington, and Wisconsin) and in Uruguay and Taiwan. Specific serial numbers for the affected units are listed in the agency's recall record.
Healthcare facilities and users should stop using the affected HLD System Model 610 units immediately. They should contact Cenorin, LLC for instructions on how to resolve the issue, which may involve repair or replacement of the temperature sensor/control system.
The recalled product
- Product
- HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.
- Manufacturer
- Cenorin, LLC
- Affected units
- 25
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 610601
- 610603
- 610703
- 610704
- 610705
- 610707
- 610708
- 610802
- 610803
- 610804
- 610901
- 610903
- 610904
- 610905
- 611001
- 611002
- 611004
- 611101
- 611102
- 611103
Distribution
Related recalls
Same manufacturer · Cenorin, LLC
See all →- HighHLD Systems Washer Pasteurizer Recalled for Terminal Block Overheating
FDA (Devices) · 2018-07-04
- HighCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- HighCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- SevereCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- ModerateCenorin HLD Systems Model 610 Washer Recalled for Cleaning Fluid Defect
FDA (Devices) · 2014-06-18
Same hazard · temperature defect
See all →- HighStryker iBur 3.0mm surgical burr recalled over thermal hazard
FDA (Devices) · 2024-09-04
Similar recalls
- HighHLD Systems Washer Pasteurizer Recalled for Terminal Block Overheating
FDA (Devices) · 2018-07-04
- ModerateCenorin HLD Systems Model 610 Washer Recalled for Cleaning Fluid Defect
FDA (Devices) · 2014-06-18
- SevereCenorin ThermaSure Medical Device Dryers Recalled for Overheating Terminal Block
FDA (Devices) · 2017-08-23
- HighDeRoyal Recalls Neonatal Temperature Skin Sensors Due to Soldering Defect
FDA (Devices) · 2015-12-09
- HighDegania Recalls Temperature Sensing Foley Catheters Due to Inspection Failures
FDA (Devices) · 2020-04-29
- SevereFDA Recalls Sorin STOCKERT Heater-Cooler Systems Due to Hydrogen Peroxide Risk
FDA (Devices) · 2017-08-02