Image intensifier detachment hazard in OEC fluoroscopy imaging systems
GE OEC field upgrade kits for fluoroscopy systems may have image intensifiers that detach during movement, potentially falling and causing injuries.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall for a risk-of-harm medical device component that may detach and cause injury. No hospitalizations or injuries have been reported to date.
Plain-English summary
GE OEC Medical Systems, Inc. is recalling Field Upgrade Kits with part numbers 5451208, 5485304, and 5761613 used with OEC Flexiview 8800, OEC 9800, and OEC 9900 medical imaging systems. These kits have been distributed worldwide, including throughout the United States.
The image intensifiers in these systems provide fluoroscopic and digital spot images. These components may become detached if the systems encounter large impact forces when moved. If the image intensifier detaches and falls, it could strike nearby personnel and cause injuries, potentially including fractures, hematomas, contusions, bruises, or pain.
The recalled product
- Product
- Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems
- Manufacturer
- GE OEC Medical Systems, Inc
- Category
- Medical Device — Imaging Systems
- Affected units
- 16
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- P/N/Sales Order Number: 5451208/80165102
- 80171955
- 80173514
- 80162453
- 80183950
- 80151733
- 80163547
- 80171532
- 80156410
- 80176650
- 10530514
- 105449333
- 80167595
- 5485304/10557644
- 118687327
- 5761613/10283014
- 10152211
- 105108003
- 105114645. OEC 9800
- UDI-DI: 00840682114349
Distribution
Part of a larger recall action
This product is one of 5 recalled by GE OEC Medical Systems, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · GE OEC Medical Systems, Inc
See all →- HighOEC 9900 Mobile Fluoroscopic X-Ray System Recall for Sealing Defect
FDA (Devices) · 2024-12-25
- HighGE OEC 9800 X-Ray System Recalled for Insufficient Tube Sealing
FDA (Devices) · 2024-12-25
- HighOEC 3D Mobile Fluoroscopic X-Ray System Tube Sealing Defect Recall
FDA (Devices) · 2024-12-25
- HighFluoroscopic X-Ray System Recalled for Insufficient X-Ray Tube Sealing
FDA (Devices) · 2024-12-25
- HighMedical X-ray Systems Image Intensifier Detachment Injury Risk
FDA (Devices) · 2023-08-09
Same hazard · equipment detachment
See all →- HighCardiohelp System HKH 8820 Wall Holder Mounting Bracket Recall
FDA (Devices) · 2025-06-04
- HighHip Traction Boot Assembly Defect Could Cause Device Detachment
FDA (Devices) · 2024-03-27
- HighAltec Welder Equipment May Detach From Aerial Device Vehicles
NHTSA · 2023-12-07
- HighNova Bus LFS Transit Buses Recalled for HVAC Unit Bracket Failure
NHTSA · 2023-11-11
- HighOEC Flexiview 8800 Image Intensifier Risks Detachment During Movement
FDA (Devices) · 2023-08-09
Similar recalls
- HighGE OEC Recalls 9800 Fluoroscopic X-Ray Systems Due to Hard Drive Defects
FDA (Devices) · 2012-08-29
- HighGE OEC 9900 Elite Mobile Fluoroscopy System Recall for Monitor Fall Risk
FDA (Devices) · 2014-04-23
- ModerateGE OEC 9800 Medical Imaging Systems Field Correction for Component Revision
FDA (Devices) · 2014-11-19
- SevereGE OEC 9800 C-arm Systems Recalled for Power Supply Defect
FDA (Devices) · 2012-08-01
- ModerateGE Healthcare Recalls FlexiView 8800 Digital Mobile Imaging Systems
FDA (Devices) · 2014-11-19
- HighGE OEC Medical Systems Recalls OEC 9800 Plus Digital Imaging Systems
FDA (Devices) · 2012-08-08