Olympus Analytical X-Ray Systems Recalled for Trigger Lock Failure
Olympus Scientific Solutions Americas is recalling Analytical X-Ray Systems because the trigger lock may fail to prevent x-ray generation in test modes.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class II and involves a defect that allows unintended ionizing radiation emission, which constitutes a significant injury risk.
Plain-English summary
Olympus Scientific Solutions Americas Corporation (OSSA) is recalling specific models of its Analytical X-Ray Systems, including the DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP-6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium, Delta Professional, and Delta-Plus. The recall involves 2,210 units distributed nationwide in the United States. The affected units are identified by having software version 2.5.17.1 or higher.
The recall is being issued because the trigger lock option does not always prevent x-ray generation. Specifically, if the unit is in test mode or repeat test mode, the trigger lock may not prevent x-ray initiation when the trigger is pulled. This defect poses a risk of unintended ionizing radiation exposure.
The correction for this issue is a software update repair. Users of these systems should contact Olympus Scientific Solutions Americas to arrange for the necessary software update to ensure the trigger lock functions correctly.
The recalled product
- Product
- Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000,
- Manufacturer
- Olympus Scientific Solutions Americas
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All Models of the Delta XRF Analyzer with Software version 2.5.17.1 and higher
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Olympus Scientific Solutions Americas
See all →- HighOlympus VANTA Analytical X-Ray System Recalled for Proximity Shutdown Failure
FDA (Devices) · 2017-08-09
- HighOlympus Vanta XRF Analyzers Recalled for LED Warning Light Failure
FDA (Devices) · 2017-02-08
- HighOlympus Vanta Analytical X-Ray System Recalled for Unintentional Emission Risk
FDA (Devices) · 2016-12-14
- ModerateOlympus Delta XRF Analyzer Recalled Due to Software Bug
FDA (Devices) · 2016-03-02
- ModerateOlympus Analytical X-ray Systems Recalled for Higher Scatter Readings
FDA (Devices) · 2015-08-12
Same hazard · unintended radiation
See all →- HighFluoroscopic X-Ray System Hand Switch May Cause Unintended Radiation
FDA (Devices) · 2024-12-11
- HighZiehm Vision RFD 3D Hand Switch Defect May Cause Unintended Radiation
FDA (Devices) · 2024-12-11
- HighX-ray system hand switch may fail, risking unintended radiation exposure
FDA (Devices) · 2024-12-11
- HighZiehm Vision RFD X-ray System Hand Switch May Cause Unintended Radiation
FDA (Devices) · 2024-12-11
- HighFluoroscopic X-ray System Hand Switch May Fail and Cause Unintended Radiation
FDA (Devices) · 2024-12-11