Olympus Analytical X-ray Systems Recalled for Higher Scatter Readings
Olympus Scientific Solutions Americas is recalling 1,047 Analytical X-ray Systems because scatter readings around the probe head were higher than published specifications for certain samples.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves inaccurate measurement readings rather than a direct physical injury or safety risk, fitting the criteria for a moderate severity recall.
Plain-English summary
Olympus Scientific Solutions Americas is recalling 1,047 Analytical X-ray Systems, including the Delta, Delta Element, and Delta Plus models, as well as specific DPO-4000 and DPO-6000 mobile units. The recall was initiated because the devices produced higher-than-published scatter readings around the probe head when testing specific samples.
The issue was identified during testing of samples known to be highly scattering, such as plastics and soils, where measurements around the probe head were higher than expected. While readings at the direct interaction point (Trigger position) were generally lower, the increased scatter around the probe head could affect the accuracy of the instrument for these specific material types. The recall covers units distributed nationwide.
Consumers and businesses using these specific models should stop using the devices and contact Olympus Scientific Solutions Americas for further instructions on how to proceed with the recall.
The recalled product
- Product
- Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10. Mobile X-ray System.
- Manufacturer
- Olympus Scientific Solutions Americas
- Hazard
- Affected units
- 1,047
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Delta
- Delta Element
- Delta Plus
- DPO-4000-XX-U-EN-EN-PM-SP12
- DPO-6000-XX-U-EN-EN-AP-SP10.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Olympus Scientific Solutions Americas
See all →- HighOlympus VANTA Analytical X-Ray System Recalled for Proximity Shutdown Failure
FDA (Devices) · 2017-08-09
- HighOlympus Vanta XRF Analyzers Recalled for LED Warning Light Failure
FDA (Devices) · 2017-02-08
- HighOlympus Vanta Analytical X-Ray System Recalled for Unintentional Emission Risk
FDA (Devices) · 2016-12-14
- SevereOlympus Analytical X-Ray Systems Recalled for Trigger Lock Failure
FDA (Devices) · 2016-08-10
- ModerateOlympus Delta XRF Analyzer Recalled Due to Software Bug
FDA (Devices) · 2016-03-02
Same hazard · inaccurate readings
See all →- HighMulti-parameter command modules may produce inaccurate cardiac output readings
FDA (Devices) · 2026-04-15
- HighBlood Oximeter Calibration Error Causes Inaccurate Oxygen Readings
FDA (Devices) · 2024-11-06
- HighSmiths Medical Arterial Kit recalled for pressure monitoring defect
FDA (Devices) · 2024-10-30
- HighGE HealthCare CRITIKON ONE-CUF blood pressure cuffs may give inaccurate readings
FDA (Devices) · 2024-06-12
- SevereCereLink ICP Extension Cable Recalled for Out of Range Readings
FDA (Devices) · 2023-12-20
Similar recalls
- SevereOlympus Analytical X-Ray Systems Recalled for Trigger Lock Failure
FDA (Devices) · 2016-08-10
- ModerateOlympus Delta XRF Analyzer Recalled Due to Software Bug
FDA (Devices) · 2016-03-02
- HighOlympus VANTA Analytical X-Ray System Recalled for Proximity Shutdown Failure
FDA (Devices) · 2017-08-09
- HighOlympus Vanta Analytical X-Ray System Recalled for Unintentional Emission Risk
FDA (Devices) · 2016-12-14
- HighOlympus Recalls High Frequency Resection Electrodes Due to Wire Breakage
FDA (Devices) · 2017-07-05
- HighGE Healthcare Recalls PET Discovery Systems Due to Scatter Overcorrection
FDA (Devices) · 2014-02-26