The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0575-2025·Announced 2024-12-11

X-ray system hand switch may fail, risking unintended radiation exposure

Three Ziehm Vision x-ray systems have hand switches that may fail under mechanical stress, potentially causing unintended radiation. Owners should verify their serial numbers and contact the manufacturer.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: This is an FDA Class II recall of a defective component that presents a risk of unintended radiation exposure. However, no illnesses or injuries have been reported. Per the severity rubric, this is classified as High (score 3) because it involves a risk-of-harm product without reported injury.

Plain-English summary

Three Ziehm Vision computed tomography x-ray systems (Serial Numbers: 9246, 90155, 90416) are subject to a recall. These devices are manufactured by Ziehm Imaging GmbH and distributed by Ziehm Imaging, Inc. The devices were distributed throughout the United States and Puerto Rico.

The wired hand switch on these systems may fail when exposed to mechanical stress such as dropping or impacts. Failure of the switching element can potentially initiate unintended radiation from the connected x-ray unit. The devices do not fully meet the requirements of IEC 60601-2-54 and 21 CFR 1020.32(c).

Facility operators should verify the serial numbers of their Ziehm Vision systems to determine if their equipment is affected. Facility staff with affected equipment should contact Ziehm Imaging, Inc. or Ziehm Imaging GmbH for guidance regarding this recall.

The recalled product

Product
Ziehm Vision system, x-ray, tomography, computed.
Manufacturer
Orthoscan, Inc.
Affected units
3

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • UDI: EZIEZIEHMVISIONFD1 Serial Numbers: 9246
  • 90155
  • 90416.

Distribution

Distributed in 1 state:

Part of a larger recall action

This product is one of 7 recalled by Orthoscan, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 7 products in this recall →

Similar recalls