The Recall Desk

Manufacturer

Orthoscan, Inc.

11 recalls in our database name Orthoscan, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
11
Critical
0
Severe
0
Most recent
2024-12-11

Of the 11 recalls from Orthoscan, Inc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–11 of 11

  • HighFDA (Devices)·Z-0579-2025·2024-12-11

    Ziehm Vision RFD 3D Hand Switch Defect May Cause Unintended Radiation

    The Ziehm Vision RFD 3D hand switch can fail when exposed to mechanical stress like dropping, potentially initiating unintended radiation from the x-ray system.

    Product
    Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0578-2025·2024-12-11

    Ziehm Vision RFD X-ray System Hand Switch May Cause Unintended Radiation

    Orthoscan's Ziehm Vision RFD fluoroscopic x-ray systems may have wired hand switches that fail under mechanical stress, potentially causing unintended radiation. The manufacturer is recalling affected units in the U.S. and Puerto Rico.

    Product
    Ziehm Vision RFD. Interventional fluoroscopic x-ray system
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0573-2025·2024-12-11

    Mobile C-arm hand switch defect may cause unintended X-ray emission

    Ziehm Solo Digital Mobile C-arm devices may have hand switch defects that can cause unintended radiation emission when exposed to impacts or drops. The recalled units have been distributed in the U.S. and Puerto Rico.

    Product
    Ziehm Solo. Digital Mobile C-arm
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0574-2025·2024-12-11

    X-ray system hand switch may fail under mechanical stress

    The Ziehm Solo FD interventional fluoroscopic x-ray system's hand switch may fail when exposed to mechanical stress, potentially causing unintended radiation. Twenty-five units are affected in the U.S. and Puerto Rico.

    Product
    Ziehm Solo FD. Interventional fluoroscopic x-ray system
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0576-2025·2024-12-11

    Fluoroscopic X-Ray System Hand Switch May Cause Unintended Radiation

    Orthoscan Inc. is recalling 22 Ziehm Vision FD fluoroscopic x-ray systems. The hand switch may fail when dropped or impacted, potentially initiating unintended radiation exposure.

    Product
    Ziehm Vision FD. Interventional fluoroscopic x-ray system
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0575-2025·2024-12-11

    X-ray system hand switch may fail, risking unintended radiation exposure

    Three Ziehm Vision x-ray systems have hand switches that may fail under mechanical stress, potentially causing unintended radiation. Owners should verify their serial numbers and contact the manufacturer.

    Product
    Ziehm Vision system, x-ray, tomography, computed.
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0577-2025·2024-12-11

    Fluoroscopic X-ray System Hand Switch May Fail and Cause Unintended Radiation

    The Ziehm Vision R fluoroscopic x-ray system has a hand switch that may fail under mechanical stress and cause unintended radiation. Ziehm Imaging, Inc. is recalling 5 affected units distributed in the U.S.

    Product
    Ziehm Vision R. Iterventional fluoroscopic x-ray system
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-0258-2018·2018-01-17

    Orthoscan Mobile Mini C-arm System Recalled for Software Defect

    Orthoscan is recalling Mobile Mini C-arm systems because a software defect allows users to activate Digital Zoom during live imaging, exposing a larger x-ray field than is visible to the user.

    Product
    Mobile Mini C-arm system Part # 1000-0001. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0259-2018·2018-01-17

    Orthoscan Mobile Mini C-arm System Recalled for Software Defect

    Orthoscan is recalling 2,012 Mobile Mini C-arm systems because a software defect allows users to activate Digital Zoom during live imaging, exposing a larger x-ray field than is visible to the user.

    Product
    Mobile Mini C-arm system Part # 1000-0004. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0261-2018·2018-01-17

    OrthoScan Recalls Mobile Mini C-arm Systems Due to Software Defect

    OrthoScan is recalling 2,012 Mobile Mini C-arm systems because a software defect allows users to activate Digital Zoom during live imaging, exposing a larger x-ray field than visible.

    Product
    Mobile Mini C-arm system Part # 1000-0004-FD. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0260-2018·2018-01-17

    OrthoScan Recalls Mobile Mini C-arm Systems Due to Software Defect

    OrthoScan is recalling 2,012 Mobile Mini C-arm systems because a software defect allows users to activate Digital Zoom during live imaging, exposing a larger x-ray field than visible.

    Product
    Mobile Mini C-arm system Part # 1000-0005. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide