Mobile X-ray System with Faulty Glue Assembly May Pose Electric Shock Hazard
Siemens Mobile X-ray systems may have incorrect glue on the tank cover assembly, which could fall off and allow liquid ingress, creating electric shock risk. The recall affects 9 units distributed in the US and internationally.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This recall describes a risk-of-harm scenario (electric shock from liquid ingress through a defective glue joint) but with no reported illnesses or injuries. The FDA Class II classification and the explicit risk of electric shock support a High severity rating. Per the rubric, a risk-of-harm product without reported injury is scored at 3.
Plain-English summary
The Cios Select VA 21 with FD-A Mobile X-ray system manufactured by Siemens Medical Solutions USA, Inc. is subject to this recall.
The tank cover assembly on these units contains a plexiglass sheet affixed with incorrect glue. The faulty adhesive may allow the plexiglass sheet to detach. If the plexiglass falls off, liquid may enter the equipment and contact internal live electrical parts, creating a risk of electric shock to users or operators who touch the contaminated area.
This recall affects 9 units that were distributed worldwide. In the United States, units were distributed to Florida, Georgia, Kansas, Minnesota, North Carolina, and New York. International distribution includes Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg, Mauritius, Mexico, Monaco, Peru, Philippines, Poland, Romania, Slovakia, South Africa, Spain, Switzerland, Taiwan, Ukraine, and the United Kingdom.
The recalled product
- Product
- Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 9
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · electrical shock
See all →- High
- Severe
- HighFort Defiance Automated Field Steam Sterilizer Recalled for Electrical Grounding Defect
FDA (Devices) · 2026-08-26
- SevereJayco Recalls 2026 Seismic Fifth Wheel Toy Haulers Due to Fire Risk
NHTSA · 2026-07-02
- Severe
Similar recalls
- HighSiemens Cios Alpha X-Ray System Recalled for Electrical Shock Risk
FDA (Devices) · 2020-06-03
- HighSiemens Recalls Cios Alpha X-Ray Systems Due to Electrical Shock Risk
FDA (Devices) · 2016-04-06
- HighSiemens Cios Alpha Mobile X-Ray System Recalled for Electrical Shock Risk
FDA (Devices) · 2020-05-20
- HighSiemens X-Ray Systems Software Issue May Cause Patient Misdiagnosis
FDA (Devices) · 2023-11-08
- HighSiemens Cios Spin Mobile X-Ray System Recalled for Electrical Shock Risk
FDA (Devices) · 2020-05-20
- SevereSiemens Cios Spin X-Ray System Recalled for Electrical Shock Risk
FDA (Devices) · 2020-06-03