The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2232-2020·Announced 2020-06-10

Zimmer Biomet Recalls Vital MIS Spinal Fixation Implants Due to Locking Issues

Zimmer Biomet is recalling Vital MIS Spinal Fixation system implants because reports indicate the extended tab screws may not lock adequately, causing the closure top to back out after surgery.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a functional defect (inadequate locking) that poses a risk of harm, but the source text does not report specific injuries or hospitalizations, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Zimmer Biomet, Inc. is recalling Vital MIS Spinal Fixation system implants, specifically the CANN BL POLY EXT 8.5X60MM (Item Number 811M8560). These are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation in skeletally mature and adolescent patients.

The recall is being conducted due to reports of inadequate locking of the extended tab screws. Complaints have been received indicating that the closure top backed out of the screw post-operatively.

A total of 3,235 units were distributed nationwide in the United States. The affected lot number is SBM111598. Healthcare providers and patients should contact Zimmer Biomet for further instructions regarding the recall and potential replacement or repair of the devices.

The recalled product

Product
Vital MIS Spinal Fixation system implants CANN BL POLY EXT 8.5X60MM, Item Number 811M8560 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is…
Affected units
3,235

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • UDI: (01)00887868326158
  • Lot Numbers: SBM111598

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 25 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 25 products in this recall →

Similar recalls