X-ray table recalled for fire hazard from power supply unit short circuit
Siemens MULTIX TOP I PRO X-ray tables are recalled due to a power supply short circuit that may cause overheating and fire. Three units have been identified.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall with a potential fire hazard, but no reported injuries, illnesses, property damage, or actual incidents. The hazard is theoretical ('may lead to') rather than confirmed, fitting the criterion for score 3: risk-of-harm products where injury has not yet been reported.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling the MULTIX TOP I PRO Radiographic X-ray Table (Material Number 475525). The device is a patient table used in radiographic imaging systems in conjunction with Siemens Multix PRO and TOP Systems devices. Three units with serial numbers 1523, 1585, and 1897 are affected.
A short circuit has been identified in the power supply unit (PSU) of the Touch Display component. This short circuit may lead to overheating and potential fire.
The affected units have been distributed nationwide in the United States. Facilities that have received these devices should contact Siemens Medical Solutions USA, Inc. to determine appropriate corrective action for units with the identified serial numbers.
The recalled product
- Product
- MULTIX TOP - Intended Use: The Multix TOP I PRO Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 475525
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 3
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI-DI: N/A Serial Numbers: 1523 1585 1897
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 9 recalled by Siemens Medical Solutions USA, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 9 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · fire
See all →- CriticalNTI and HTP Gas Boilers Recalled for Fire and Carbon Monoxide Risks
CPSC · 2026-10-29
- SevereWhatever Works Recalls Microwave Grills Due to Fire Hazard
CPSC · 2026-10-29
- SevereHAIARA Power Banks Recalled Due to Fire and Burn Hazards
CPSC · 2026-10-29
- SevereDREO Recalls DR-HSH015 Space Heaters Due to Fire and Burn Hazards
CPSC · 2026-10-29
- SeverePamag Pieeoo Mattresses Recalled for Fire Hazard Risk
CPSC · 2026-10-29
Similar recalls
- SevereSiemens AXIOM Multix MT Patient Table Recalled for Cable Failure Risk
FDA (Devices) · 2018-11-07
- HighSiemens Recalls Multix Fusion X-Ray Systems Due to Table Tilt Risk
FDA (Devices) · 2016-10-12
- HighSiemens Recalls Multix Fusion VA10 Radiographic Systems Due to Table Defect
FDA (Devices) · 2016-10-12
- HighMedical X-ray Systems Recalled for Electrical Fire Hazard
FDA (Devices) · 2023-04-19
- SevereSiemens Multix Fusion Radiographic Systems Recalled for Lifting Column Cable Defect
FDA (Devices) · 2018-11-07
- SevereSiemens Uroskop Access Fluoroscopic X-Ray System Recalled for Table Fall Hazard
FDA (Devices) · 2012-09-19