GE Healthcare Recalls Tec 6 Anesthesia Vaporizers Due to Leak Detection Issue
GE Healthcare is recalling Tec 6 and Tec 6 Plus anesthesia vaporizers because the pre-operative leak test may fail to detect leaks caused by seal wear.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. Per the rubric, risk-of-harm products where injury has not yet been reported are assigned a score of 3 (High).
Plain-English summary
GE Healthcare, LLC is recalling the Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The recall involves 31,443 devices distributed worldwide, including 23,947 units in the United States and 7,496 units internationally. These electronic vaporizers are used to deliver the anesthetic agent desflurane and attach to specific Datex-Ohmeda and NAD anesthesia machines.
The recall was initiated because the pre-operative check, specifically the low pressure leak test at the 1% dial setting, may not detect the full range of leaks resulting from seal wear degradation in the vaporizers. If a leak is not detected during this test, it will result in fresh gas leaking from the device.
The source text does not specify the corrective actions consumers or healthcare providers should take, nor does it report any specific illnesses or injuries associated with this defect.
The recalled product
- Product
- Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthetic agent desflurane. The Tec 6 Plus attaches only to Datex-Ohmeda anesthesia machines equipped…
- Manufacturer
- GE Healthcare, LLC
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Part Numbers: X1107-9001-000
- X1107-9002-000
- X1107-9003-000
- X1107-9004-000
- X1107-9005-000
- X1107-9006-000
- X1107-9007-000
- X1109-9008-000
- X1109-9009-000
- X1107-9010-000
- X1107-9011-000
- X1107-9012-000
- X1107-9014-000
- X1107-9015-000
- X1107-9016-000
- X1107-9101-000
- X1107-9102-000
- X1107-9103-000
- X1107-9201-000
- X1107-9202-000
Distribution
Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Healthcare CT, PET, and NM Patient Tables May Drop During Service
FDA (Devices) · 2022-08-17
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- SevereGE Healthcare Anesthesia Systems Recalled for Potential Transducer Connection Reversal
FDA (Devices) · 2022-07-27
Same hazard · leak
See all →- SevereFresenius Medical Care Recalls 5008X Standard HD Pre-Flush Blood Tubing Sets
FDA (Devices) · 2026-09-09
- HighAVID Medical Recalls Halyard LAVH PACK Due to Sterilization Error
FDA (Devices) · 2026-08-26
- HighMedline Convenience Kits with Recalled Swan-Ganz Catheters
FDA (Devices) · 2026-07-08
- HighBicarby dialysate luer-lock connections may leak fluid onto floors
FDA (Devices) · 2026-06-17
- HighSwan-Ganz Pacing Catheter Model D200F7 Recall Due to Leak Risk
FDA (Devices) · 2026-05-27
Similar recalls
- ModerateGE Healthcare Recalls 28 Datex-Ohmeda TEC 7 Anesthesia Vaporizers
FDA (Devices) · 2018-04-04
- HighGE Tec Anesthesia Vaporizers Recalled for Insufficient Anesthetic Delivery
FDA (Devices) · 2025-01-15
- HighAnesthesia Vaporizers May Deliver Less Anesthetic Than Dial Setting Due to Manufacturing Defect
FDA (Devices) · 2025-01-15
- HighTec 820 ISO Anesthesia Vaporizers can deliver less anesthetic than dial setting
FDA (Devices) · 2025-01-15
- HighTec 820 SEV Anesthesia Vaporizer Recalled for Inadequate Anesthetic Delivery
FDA (Devices) · 2025-01-15
- HighTec 850 SEV Anesthesia Vaporizers Recalled for Insufficient Anesthetic Delivery
FDA (Devices) · 2025-01-15