Proteus 235-Proton Therapy System May Deliver Unintended Radiation
The Proteus 235-Proton Therapy System may deliver proton therapy without prescribed beam gating if the user forgets to select the trigger input. This could cause radiation to be positioned incorrectly.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II FDA recall of a risk-of-harm medical device with no reported injuries or illnesses. The recall addresses a potential for incorrect radiation delivery due to beam gating omission, creating risk of mispositioned treatment fields, but no adverse events have been documented.
Plain-English summary
The Proteus 235-Proton Therapy System is a medical device used to deliver targeted proton therapy. It can be configured to use beam gating, a safety mechanism that restricts when radiation is delivered during treatment.
If a treatment operator forgets to manually select the trigger input when beam gating has been prescribed, the system will deliver proton radiation without the prescribed gating. When the beam is not gated as intended, the treatment field may be positioned incorrectly, potentially causing radiation to reach unintended areas.
Hospitals and treatment centers using this system should review their treatment protocols to ensure beam gating selection is confirmed when prescribed. Safety procedures should include verification steps to ensure that beam gating is properly activated before treatment begins.
The recalled product
- Product
- Proteus 235-Proton Therapy System
- Manufacturer
- Ion Beam Applications S.A.
- Category
- Medical Device — Proton Therapy
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (13)
- Serial Numbers: PAT.107 (EU)
- PAT.108 (US)
- PAT.109 (US)
- PAT.115 (EU)
- SAT.117 (EU)
- SAT.118 (RU)
- SAT.119 (US)
- SAT.120 (EU)
- SAT.122 (EU)
- SAT.125 (IN)
- SAT.126 (US)
- SAT.132 (EU)
- SAT.133 (US).
Distribution
Related recalls
Same manufacturer · Ion Beam Applications S.A.
See all →- HighIBA proton therapy system may not interrupt beam when gating fails
FDA (Devices) · 2026-03-18
- HighProton Therapy System Software Allows Safety Verification Mechanisms to Be Disabled
FDA (Devices) · 2024-09-18
- HighIBA Proton Therapy System PROTEUS 235 Collision Detection Malfunction
FDA (Devices) · 2024-08-28
- HighProton Therapy System May Continue Irradiation When Accelerator Control Parameters Out of Tolerance
FDA (Devices) · 2024-06-12
- HighProton Therapy System Recall: Safety Parameter Failures Risk Unintended Radiation
FDA (Devices) · 2024-05-01
Same hazard · treatment positioning error
See all →- HighProton therapy system may display insufficient CBCT images during positioning
FDA (Devices) · 2024-09-18
- HighZAP-X Radiosurgery System Software Issue May Cause Treatment Misalignment
FDA (Devices) · 2022-09-21
Similar recalls
- HighIBA proton therapy system may not interrupt beam when gating fails
FDA (Devices) · 2026-03-18
- HighProton Therapy System Delivers Incorrect Radiation Dose in First Layer
FDA (Devices) · 2022-06-08
- HighProteus 235 Proton Therapy System May Display Misleading Popup Message
FDA (Devices) · 2021-09-15
- HighProteus 235 Proton Beam Device Recalled for Correction Vector Loss
FDA (Devices) · 2018-07-18
- HighProton Therapy System May Continue Irradiation When Accelerator Control Parameters Out of Tolerance
FDA (Devices) · 2024-06-12
- HighProton Therapy Systems May Fail to Verify Beam Range After Pause
FDA (Devices) · 2022-03-16