The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1742-2022·Announced 2022-09-21

ZAP-X Radiosurgery System Software Issue May Cause Treatment Misalignment

A software issue in the ZAP-X Radiosurgery System may cause incorrect treatment positioning in patients with large rotational deviations. The alignment algorithm could miscalculate treatment table offset values, potentially affecting treatment delivery.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: FDA Class II device with potential for treatment positioning errors. No illnesses or injuries have been reported to date, making the maximum applicable score 3 per the rubric's criteria for risk-of-harm products without reported injury.

Plain-English summary

The ZAP-X Radiosurgery System is a medical device used for radiation treatment. A software issue affecting initial patient setup has been identified and recalled.

The issue is related to the transitional alignment algorithm, which calculates treatment table positions. When patients have large rotational deviations (1.5 degrees or greater) between digitally reconstructed radiographs and pre-delivery X-ray images, this algorithm may incorrectly calculate the treatment table offset values. This results in an incorrect position for the treatment isocenter.

Eight systems are affected. In the United States, affected systems are located in Arizona, Maryland, Florida, and Colorado. Internationally, affected systems are located in Germany, Switzerland, and Japan. Affected systems are identified by specific serial numbers and use Treatment Delivery Software Version 1.8.53.

The recalled product

Product
ZAP-X Radiosurgery System
Affected units
8

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (11)

  • UDI-DI: 00860183001504
  • Serial Numbers
  • ZUC17152
  • ZUC17153
  • ZUC18155
  • ZUC20157
  • ZUC20158
  • ZUC20156
  • ZUC21161
  • ZUC21162
  • using Treatment Delivery Software Version: 1.8.53

Distribution

Distributed nationwide across the United States.

The notice also names 4 states specifically:

Same manufacturer · Zap Surgical Systems